Specifications
Measurements
Available measurements
V
RMS
- Volts RMS
VDF - Volts Distortion Factor
A
RMS
- Amps RMS
VTIF - Volts Telephone Influence Factor
WATT - Watts
ATHD - Amps Total Harmonic Distortion
VA - Volt-Amps
ADF - Amps Distortion Factor
VAR - Volt-Amps Reactive
ATIF - Amps Telephone Influence Factor
FRQ - Frequency
VF - Fundamental Volts rms
PF - Power factor
AF - Fundamental Amps rms
VPK+ - Volts peak (positive)
IMP - Impedance
VPK- - Volts peak (negative)
RES - Resistance
APK+ - Amps peak (positive)
REA - Reactance
APK- - Amps peak (negative)
HR - Integrator time
VDC - Volts DC
WHR - Watt Hours
ADC - Amps DC
VAH - VA Hours
VRMN - Volts rectified mean
VRH - VAr Hours
ARMN - Amps rectified mean
AHR - Amp Hours
VCF - Voltage crest factor
WAV - Average Watts
ACF - Current crest factor
PFAV - Average Power Factor
VTHD - Volts Total Harmonic Distortion
CVAR - Correction VArs
Voltage and current ranges
Voltage Ranges
2000 V
peak
, 1000 V
peak
, 500 V
peak
, 200 V
peak
, 100 V
peak
, 50 V
peak
, 20 V
peak
, 5 V
peak
, 2 V
peak
Current Ranges (30 A Shunt)
200 A
peak
, 100 A
peak
, 50 A
peak
, 20 A
peak
, 10 A
peak
, 5 A
peak
, 2 A
peak
, 1 A
peak
, 0.5 A
peak
, 0.2 A
peak
, 0.1 A
peak
Current Ranges (1 A Shunt)
5 A
peak
, 2 A
peak
, 1 A
peak
, 0.5 A
peak
, 0.2 A
peak
, 0.1 A
peak
, 0.25 A
peak
, 0.0125 A
peak
, 0.005 A
peak
, 0.0025 A
peak
Current Ranges (External
Shunt)
3 V
peak
, 1.5 V
peak
, 0.75 V
peak
, 0.3 V
peak
, 0.15 V
peak
, 0.075 V
peak
, 0.03 V
peak
, 0.015 V
peak
, 0.0075 V
peak
Measurement accuracy - voltage
and current
Voltage Accuracy, V
RMS
( 45 Hz - 850 Hz )
± 0.04% of Reading ± 0.04% of Range ± 0.02 V
Voltage Accuracy, V
RMS
( 10 Hz - 45 Hz, 850 Hz -
1 MHz )
± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± 0.02 V (typical)
Voltage Accuracy, DC
± 0.05% of Reading ± 0.1% of Range ± 0.05 V
Voltage Accuracy, V
rmn
± 0.2% of Reading ± 0.1% of Range ± 0.1 V
Effect of Common Mode
(Typical)
1000 V, 60 Hz < 10 mV
100 V, 100 kHz < 50 mV
Current Accuracy, A
RMS
( 45 Hz - 850 Hz )
± 0.04% of Reading ± 0.04% of Range ± (20 µV/Z
ext
)
Current Accuracy, A
RMS
( 10 Hz - 45 Hz, 850 Hz -
1 MHz )
± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± (20 µV/Z
ext
) (typical)
Current Accuracy, DC
± 0.05% of Reading ± 0.1% of Range ± (50 µV/Z
ext
)
Current Accuracy, A
RMS
± 0.2% of Reading ± 0.1% of Range ± (100 µV/Z
ext
)
Datasheet