Specifications

Measurements

Available measurements

V

RMS

- Volts RMS

VDF - Volts Distortion Factor

A

RMS

- Amps RMS

VTIF - Volts Telephone Influence Factor

WATT - Watts

ATHD - Amps Total Harmonic Distortion

VA - Volt-Amps

ADF - Amps Distortion Factor

VAR - Volt-Amps Reactive

ATIF - Amps Telephone Influence Factor

FRQ - Frequency

VF - Fundamental Volts rms

PF - Power factor

AF - Fundamental Amps rms

VPK+ - Volts peak (positive)

IMP - Impedance

VPK- - Volts peak (negative)

RES - Resistance

APK+ - Amps peak (positive)

REA - Reactance

APK- - Amps peak (negative)

HR - Integrator time

VDC - Volts DC

WHR - Watt Hours

ADC - Amps DC

VAH - VA Hours

VRMN - Volts rectified mean

VRH - VAr Hours

ARMN - Amps rectified mean

AHR - Amp Hours

VCF - Voltage crest factor

WAV - Average Watts

ACF - Current crest factor

PFAV - Average Power Factor

VTHD - Volts Total Harmonic Distortion

CVAR - Correction VArs

Voltage and current ranges

Voltage Ranges

2000 V

peak

, 1000 V

peak

, 500 V

peak

, 200 V

peak

, 100 V

peak

, 50 V

peak

, 20 V

peak

, 5 V

peak

, 2 V

peak

Current Ranges (30 A Shunt)

200 A

peak

, 100 A

peak

, 50 A

peak

, 20 A

peak

, 10 A

peak

, 5 A

peak

, 2 A

peak

, 1 A

peak

, 0.5 A

peak

, 0.2 A

peak

, 0.1 A

peak

Current Ranges (1 A Shunt)

5 A

peak

, 2 A

peak

, 1 A

peak

, 0.5 A

peak

, 0.2 A

peak

, 0.1 A

peak

, 0.25 A

peak

, 0.0125 A

peak

, 0.005 A

peak

, 0.0025 A

peak

Current Ranges (External
Shunt)

3 V

peak

, 1.5 V

peak

, 0.75 V

peak

, 0.3 V

peak

, 0.15 V

peak

, 0.075 V

peak

, 0.03 V

peak

, 0.015 V

peak

, 0.0075 V

peak

Measurement accuracy - voltage
and current

Voltage Accuracy, V

RMS

( 45 Hz - 850 Hz )

± 0.04% of Reading ± 0.04% of Range ± 0.02 V

Voltage Accuracy, V

RMS

( 10 Hz - 45 Hz, 850 Hz -
1 MHz )

± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± 0.02 V (typical)

Voltage Accuracy, DC

± 0.05% of Reading ± 0.1% of Range ± 0.05 V

Voltage Accuracy, V

rmn

± 0.2% of Reading ± 0.1% of Range ± 0.1 V

Effect of Common Mode
(Typical)

1000 V, 60 Hz < 10 mV

100 V, 100 kHz < 50 mV

Current Accuracy, A

RMS

( 45 Hz - 850 Hz )

± 0.04% of Reading ± 0.04% of Range ± (20 µV/Z

ext

)

Current Accuracy, A

RMS

( 10 Hz - 45 Hz, 850 Hz -
1 MHz )

± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± (20 µV/Z

ext

) (typical)

Current Accuracy, DC

± 0.05% of Reading ± 0.1% of Range ± (50 µV/Z

ext

)

Current Accuracy, A

RMS

± 0.2% of Reading ± 0.1% of Range ± (100 µV/Z

ext

)

Datasheet

4 www.tektronix.com