beautypg.com

High voltage differential probes, Characteristics – Atec Tektronix-P5205-P5210 User Manual

Page 2

background image

High Voltage Differential Probes

P5200 • P5205 • P5210

High Voltage Differential Probes • www.tektronix.com/accessories

2

overrange visual and audible indicator

which warns the user when they are

exceeding the linear range of the probe. It

can be used with Tektronix TEKPROBE

interface oscilloscopes directly or with any

oscilloscope with the use of the 1103

TEKPROBE

®

Power Supply.

The P5205 is a 100 MHz Active Differential

Probe capable of measuring fast rise times

of signals in floating circuits. This 1,300 V

differential probe can measure safely

voltages in IGBT circuits such as motor

drives or power converters. It is specifically

designed to operate on Tektronix oscillo-

scopes with TEKPROBE interface.

Characteristics

Specifications

P5200

P5205

P5210

Bandwidth (–3 dB)

25 MHz

100 MHz

50 MHz

Differential Mode Voltage

1000 V CAT II

1000 V CAT II

4400 V

(RMS or DC) (between positive
and negative input leads)

Differential Mode Voltage

1300 V

1300 V

5600 V

(Peak)

*1

(between positive

and negative input leads)

Common Mode Voltage (RMS or DC

1000 V CAT II

1000 V CAT II

2200 V CAT II

+ Pk AC) (between positive lead and

600 V CAT III

600 V CAT III

1000 V CAT III

ground or negative lead and ground)

Attenuation (switchable)

50X/500X

50X/500X

100X/1000X

CMRR at 60 Hz/1 MHz (typical)

80 dB/50 dB

80 dB/50 dB

80 dB/50 dB

Input R (each input)

4 M

Ω

4 M

Ω

8 M

Ω

Input C (each input)

7 pF

7 pF

7 pF

Power Source

AC wall adapter

TEKPROBE Power Supply

TEKPROBE Power Supply

Safety (all)

UL3111-1/CSA1010.1/CSA1010.2.031/IEC61010-2-031/EN61010-2-031

Compatible Products

All

*3

TDS400/500/600/700/ TDS400/500/600/700/

TDS3000B/5000B/6000B

*2

/7000B

*2

/

TDS3000B/5000B/6000B

*2

/7000B

*2

/

CSA7000B

*2

and 11000 Series

CSA7000B

*2

and 11000 Series

*1

The RMS value of the waveform shall not exceed the Common Mode Voltage (RMS) for either lead.

*2

TDS7154B, TDS7254B, TDS7404B, TDS6404, TDS6604, CSA7154, and CSA7404B require a TCA-1MEG Buffer Amplifier.

*3

WARNING: For safe operation, do not use the P5200 High Voltage Differential Probe with oscilloscopes that have floating inputs (isolated inputs), such as the Tektronix TPS2000 series oscilloscopes
and THS700 series oscilloscopes. The P5200 High Voltage Differential Probe requires an oscilloscope or other measurement instrument with grounded inputs.