High voltage differential probes, Characteristics – Atec Tektronix-P5205-P5210 User Manual
Page 2
![background image](/manuals/319898/2/background.png)
High Voltage Differential Probes
P5200 • P5205 • P5210
High Voltage Differential Probes • www.tektronix.com/accessories
2
overrange visual and audible indicator
which warns the user when they are
exceeding the linear range of the probe. It
can be used with Tektronix TEKPROBE
™
interface oscilloscopes directly or with any
oscilloscope with the use of the 1103
TEKPROBE
®
Power Supply.
The P5205 is a 100 MHz Active Differential
Probe capable of measuring fast rise times
of signals in floating circuits. This 1,300 V
differential probe can measure safely
voltages in IGBT circuits such as motor
drives or power converters. It is specifically
designed to operate on Tektronix oscillo-
scopes with TEKPROBE interface.
Characteristics
Specifications
P5200
P5205
P5210
Bandwidth (–3 dB)
25 MHz
100 MHz
50 MHz
Differential Mode Voltage
1000 V CAT II
1000 V CAT II
4400 V
(RMS or DC) (between positive
and negative input leads)
Differential Mode Voltage
1300 V
1300 V
5600 V
(Peak)
*1
(between positive
and negative input leads)
Common Mode Voltage (RMS or DC
1000 V CAT II
1000 V CAT II
2200 V CAT II
+ Pk AC) (between positive lead and
600 V CAT III
600 V CAT III
1000 V CAT III
ground or negative lead and ground)
Attenuation (switchable)
50X/500X
50X/500X
100X/1000X
CMRR at 60 Hz/1 MHz (typical)
80 dB/50 dB
80 dB/50 dB
80 dB/50 dB
Input R (each input)
4 M
Ω
4 M
Ω
8 M
Ω
Input C (each input)
7 pF
7 pF
7 pF
Power Source
AC wall adapter
TEKPROBE Power Supply
TEKPROBE Power Supply
Safety (all)
UL3111-1/CSA1010.1/CSA1010.2.031/IEC61010-2-031/EN61010-2-031
Compatible Products
All
*3
TDS400/500/600/700/ TDS400/500/600/700/
TDS3000B/5000B/6000B
*2
/7000B
*2
/
TDS3000B/5000B/6000B
*2
/7000B
*2
/
CSA7000B
*2
and 11000 Series
CSA7000B
*2
and 11000 Series
*1
The RMS value of the waveform shall not exceed the Common Mode Voltage (RMS) for either lead.
*2
TDS7154B, TDS7254B, TDS7404B, TDS6404, TDS6604, CSA7154, and CSA7404B require a TCA-1MEG Buffer Amplifier.
*3
WARNING: For safe operation, do not use the P5200 High Voltage Differential Probe with oscilloscopes that have floating inputs (isolated inputs), such as the Tektronix TPS2000 series oscilloscopes
and THS700 series oscilloscopes. The P5200 High Voltage Differential Probe requires an oscilloscope or other measurement instrument with grounded inputs.