beautypg.com

Atec Tenney-ETCU-16 User Manual

Page 2

background image

Model

ETCU–09

ETCU–16

ETCU–30

ETCU–64

Interior Volume

9 cu. ft.

16 cu. ft.

30 cu. ft.

64 cu. ft.

256 liters

460 liters

850 liters

1,812 liters

Interior Dimensions

24 x 25 x 26

36 x 26 x 30

40 x 36 x 36

48 x 48 x 48

WxDxH - inches (mm)

(610 x 635 x 660)

(914 x 660 x 762)

(1016 x 914 x 914)

(1219 x 1219 x 1219)

Exterior Dimensions

34 x 73 x 78

46 x 74 x 82

50 x 87 x 88

58 x 99 x 100

WxDxH - inches (mm)

(864 x 1854 x 1983)

(1169 x 1881 x 2084)

(1271 x 2211 x 2237)

(1474 x 2516 x 2542)

Nominal Horsepower

3.5 x 3.5

6 x 6

10 x 10

3.5 x 3.5

6 x 6

10 x 10

3.5 x 3.5

6 x 6

10 x 10 15 x 15

3.5 x 3.5

6 x 6

10 x 10 15 x 15

Rated Horsepower

2.5 x 2.5

5 x 5

10 x 10

2.5 x 2.5

5 x 5

10 x 10

2.5 x 2.5

5 x 5

10 x 10 15 x 15

2.5 x 2.5

5 x 5

10 x 10 15 x 15

Compressor Type

Scroll

Scroll

Scroll

Scroll

Noise (dBa)

Heating

60

60

60

60

60

60

60

60

60

60

60

60

60

60

Cooling

68

75

75

68

75

75

68

75

75

75

68

75

75

75

Heating
Elements (Kw)

230/460

6

12

12

6

12

12

6

12

12

24

6

12

12

24

208V

4.5

9

9

4.5

9

9

4.5

9

9

18

4.5

9

9

18

Model

ETCU–09

ETCU–16

ETCU–30

ETCU–64

Cooling Rate (min)

190

°

C to -65

°

C

49

34

14

78

36

19

155

76

31

20

256

125

52

42

71

°

C to -65

°

C

36

26

11

40

28

15

119

59

26

17

197

97

44

33

85

°

C to -40

°

C

27

18

8

32

20

12

95

42

21

12

157

69

36

24

Heating Rate (min)

-65

°

C to 190

°

C

35

16

16

32

22

22

60

30

30

20

110

53

53

27

-65

°

C to 71

°

C

17

8

8

14

11

11

27

13

13

10

45

23

23

13

-40

°

C to 85

°

C

16

8

8

12

10

10

27

13

13

11

45

23

23

14

Water Cooled

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Air Cooled

Y

Y

Y

Y

Y

Y

Y

208/230V-1 Ph

Y

N/A

N/A

Y

N/A

N/A

Y

N/A

N/A

N/A

Y

N/A

N/A

N/A

208/230V-3 Ph

Y

Y

N/A

Y

Y

N/A

Y

Y

N/A

N/A

Y

Y

N/A

N/A

460V-3 Ph

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Air Delivery
(Blower HP)

.5 HP

(2) .5 HP

(2) .75 HP

(2) .75 HP

Airfl ow (CFM)

750

1600

1600

1600

Temp Unif. (Std
Dev/9 Pt)

+/-0.5

°

C/+/-1

°

C

+/-0.5

°

C/+/-1

°

C

+/-0.5

°

C/+/-1

°

C

+/-0.5

°

C/+/-1

°

C

Temp Control

+/-0.3

°

C

+/-0.3

°

C

+/-0.3

°

C

+/-0.3

°

C

* Single Stage Refrigeration Temperature Range: -35°C to 200°C
* Performance is based upon an empty chamber operatfing at 24°C (75°F) ambient air and may vary slightly at other ambient temperatures. Voltages

below those listed may affect performance. All performance data is for 240 or 480V / 60Hz operation.

* Liquid Nitrogen is available as an option on all systems. When Liquid Nitrogen is provided, the heater KW is automatically upgraded to the maximum

for that chamber size. Consult factory for current draw of any Liquid Nitrogen equipped chambers.

* Airflow circulation is based on blower wheel performance curves. Actual circulation rate will vary depending on the cooling option selected.
* Temperature uniformity Standard deviation from the mean, measured at –25°C and 100°C. 9 Point: Uniformity measured in accordance with ASTM

E145 section 4 at temperatures of –25°C and 100°C.

* Test chamber performance may be affected by the addition of certain optional accessories.
* Chamber operation utilizing 50Hz power utility will derate performance approximately 17%
* It is Thermal Product Solutions' policy to constantly improve quality, features and performance. Thermal Product Solutions reserves the right to

change specifications without notice.

Gruenberg, Blue M, Tenney
Mailing Address: P.O. Box 150 | White Deer, PA 17887-0150 | USA |
Phone: (570) 538-7200 | Fax: (570) 538-7380
Physical Address: 2821 Old Route 15 | New Columbia, PA 17856-9396
[email protected]

Specifications & Product Information is subject to change without notice.