Atec Temptronic-TP04100A User Manual
Syst em specifica tions, Dimensions and weight, Options and accessories
Model
TP04
1
00A
ThermoStr
eam
®
S
yst
em Specifica
tions
TP04100A
Dimensions and Weight
Dimensions:
Controller: 40.1 cm (15.8 inches) wide x 14.8 cm (5.8 inches) high x 48.9 cm (19.2 inches) deep
Thermal Wand:
8 cm (3 inches) in diameter x 36 cm (14 inches) long
Stand:
48.25 cm (19 inches) approximately above tabletop to top of post
Reach from vertical post: Standard: 51 cm (20 inches)
Weight:
Controller with Thermal Wand and Hose ................................................. 14.5 kg (32 lbs)
Thermal Wand Stand Assembly ................................................................ 7 kg (15.5 lbs)
Thermal Cap ........................................................................................... .45 kg (1 lb)
Options and Accessories
Thermal Wand Stand Assembly
Expanding, locking stand assembly rotates 340° about the controller cabinet for
positioning the Thermal Wand directly over the DUT.
Benchmount Stand Assembly
Attaches the Thermal Wand Stand Assembly directly to a bench. Can be mounted up to
30.5 cm (12 inches) from controller chassis.
Thermal
Cap
Transparent single wall glass cap of approximately 9.9 cm (3.9 inches)
ID (Internal Diameter); includes metal guard.
Shroud Kit
Shroud attaches to thermal wand for coupling to devices. Kit contains a variety of sizes
and shapes (See price list). Available in conductive and non-conductive material.
Insulation Kit
Non-conductive silicone foam rubber sheets.
Three sheets: (12” x 12” x .125”), (12” x 12” x .25”), (12” x 12” x .5”)
Custom Thermal Test Enclosures
Frost-free thermal enclosures for testing ICs, hybrids, modules and small subassemblies
and ThermoFixture
®
precise temperature in a moisture-free thermal environment. Available in standard and
custom configurations. ThermoFixture can include ATE interface.
Pneumatic Wand Lift Assembly
Automates the raising and lowering of the Thermal Wand for the loading and unloading of
the DUT.
Temperature Range
1
-20° to +225°C
Temperature Display and Resolution
0.1°C
Temperature Accuracy
±1°C
Facilities Preparation
Power Requirements
1
115V AC ±10%, 50 or 60 Hz, 1 ø, 15 amp
230V AC ±10%, 50 or 60 Hz, 1 ø, 10 amp
Compressed Air Supply Pressure
2
5.6 to 7 kg /cm
2
(80 to 90 PSIG),
5.6 kg/cm
2
(80 PSIG) nominal
Supply Airflow at
4 to 5 liters/second (8-10 scfm)
Nominal Supply Pressure
2
4.5 l/s (9 scfm) nominal
Dewpoint
3
ambient to -40°C
Air Supply Temperature
+20° to +28°C (+22°C nominal
1
)
Operating Temperature (Ambient)
+20° to +28°C (+22°C nominal
1
)
Humidity
0 to 60% (45% nominal
1
)
Noise Level
≤ 70 dBA
Power Connection
Standard male IEC320 power connector socket for
attachment of IEC320 power cord (meets IEC requirements).
Power cord (2.4 meters in length) included.
1
Reduced performance may be encountered at operating conditions less than or greater than nominal.
2
Requires clean, dry air supply.
3
-40°C dewpoint is nominal for low temperature testing for extended periods.
(When calibrated against
NIST Transfer Standard)