Atec Zimmer-LMG450 User Manual
Page 7
3 Phase/
4 Wire
3 Phase/
3 Wire
Line to Line
Voltage
Line to
Neutral/Earth
66/115V
115V
66V
120V
120V
69V
120/208V
208V
120V
240V
240V
139V
230/400V
400V
230V
277/480V
480V
277V
500V
500V
289V
400/690V
690V
400V
1000V
1000V
578V
Voltage measuring ranges
Nominal value /V
6
12.5
25
60
130
250
400
600
Maximum trms value /V
7.2
14.4
30
60
130
270
560
720
Maximum peak value for full scale /V
12.5 25
50
100
200
400
800
1600
Overload capability
1500V for 1s
Input impedance
1MΩ, 20pF
Current measuring ranges
Nominal value /A
0.6
1.2
2.5
5
10
16
Maximum trms value /A
1.3
2.6
5.2
10
18
18
Maximum peak value for full scale /A
1.875 3.75
7.5
15
30
60
Overload capability
18A permanent, 50A for 1s, 150A for 20ms
Input impedance
2mΩ
Isolation
All direct current and voltage inputs of power measuring channels against each other and against earth isolated, max. 600V/CATIII
Voltage measuring ranges for external
isolated current transduceers
Nominal value /V
0.12 0.25
0.5
1
2
4
Maximum trms value /V
0.15 0.3
0.6
1.2
2.5
5
Maximum peak value for full scale /V
0.25 0.5
1
2
4
8
Overload capability
250V for 1s
Input impedance
100kΩ, 10pF
Measuring range selection
Automatic, manual or remotely controlled
Measuring accuracy
Accuracies based on:
1. sinusoidal voltage and current
4. definition of power range as the product of
2. ambient temperature 23 °C
current and voltage range, 0 ≤ IλI ≤ 1, (λ=Power factor=P/S)
3. warm up time 1h
5. calibration interval 12 month
Other values
All other values are derived from the current, voltage and active power values. Accuracies for derived values depend on the
functional relationship (e.g. S = I * U, DS/S = DI/I + DU/U)
Synchronization
The measurement is synchronized on the signals period. There is a choice to determine the period from u(t), i(t), p(t), further
u
2
(t), i
2
(t) by using a settable filter . By this very stable readings are achieved, even at signals of pulse width modulated
frequency inverter and amplitude modulated electronic ballast, synchronization also with external signal or „Line“
Scope function
Graphical representation of sampled values over the time
Plot function
Time diagram of max. four readings, minimal resolution 50ms
Harmonic analysis prCE Harm
Measuring of current and voltage according to EN61000-4-7 with evaluation according to EN61000-3-2 (Pre-compliance)
Harmonic analysis Harm100
Analysis of current, voltage and power up to 99
th
harmonics (max. 10kHz), in total 100 harmonics, when including DC part.
Fundamental in the range from 1Hz to 1.2kHz. By selectable integer divider (1...50) a new reference fundamental can be created
as to detect interharmonics.
Flicker measuring
Flicker Meter according to EN61000-4-15 with evaluation according to EN61000-3-3
Transients – monitoring and storing
Storing and graphical displaying of transients with a resolution of 20µs. Storing depth is 1.4 Millions sample values/channel,
selectable recording duration from 0.05 to 60 seconds. Adjustable pre-trigger, different possibilities of triggering
Computer interfaces
Interfaces:
RS232 and IEEE488.2, only one interface can be used at the same time
Remote control
All functions can be remote controlled, keyboard lock for measuring parameters available
Output data
Output of all readable data, data formats equal for all interface types, SCPI command set
Transfer rate
RS232: max.115200 Baud, IEEE488.2: max. 1MByte/sec
Printer interface
Parallel PC-Printer interface with 25-pin SUB-D socket for printing measuring values, tables, graphics to matrix, inkjet or laser printers
Processing signal interface
25 pin SUB-D socket with (The option processing signal interface can be built in twice):
• four analog inputs for registration of process magnitudes (16Bit, ±10V, 1kHz)
• four analog outputs for output of readings or measured magnitudes (16Bit, ±10V, 100kHz)
• four digital inputs for registration of states (1kHz, U
LOW
<1V, U
HIGH
=4...60V/2.5mA)
• four digital outputs to signal states and alarms (open collector, output high max. 30V@100µA, output low max. 1.5V@100mA)
• one input for registration of frequency (0.1Hz...500kHz) and rotation direction of motors (U
LOW
<1V, U
HIGH
=4...10V, 1MΩ)
In- and outputs are isolated groupwise against each other and against the other electronics (testing voltage 500V)
Technical Data
Dimensioning of insulation
for all standard low voltages
The measurement inputs are
dimensioned for 600V/CAT III,
with option L45-O15 up to
1000V. This makes it possible to
measure in all standard 3-phase
low voltage networks.
The adjacent table shows that
the voltage “Line to Neutral/
Earth” is always less than 600V.
Measuring accuracy
± (% of measuring value + % of measuring range)
DC
1Hz..1kHz
45...65Hz, AC-Coupling
1kHz..5kHz
5kHz...20kHz
Voltage
0.2+0.2
0.1+0.1
0.05+0.05
0.2+0.2
0.3+0.4
Current (direct)
0.4+0.4
0.15+0.1
0.05+0.05
0.2+0.2
0.5+0.5
Active power (direct)
0.5+0.5
0.2+0.1
0.07+0.08
0.3+0.2
0.6+0.5
Current (via ext. current transducer)
0.2+0.2
0.1+0.1
0.05+0.05
0.2+0.2
0.3+0.4
Active power (via ext. current transducer)
0.3+0.3
0.15+0.1
0.07+0.08
0.3+0.2
0.6+0.5