Atec Teseq-Mil-Std416F-CS114 User Manual
Page 4
Teseq GmbH
Landsberger Str. 255 12623 Berlin Germany
T + 49 30 56 59 88 35 F + 49 30 56 59 88 34
[email protected]
www.teseq.com
4
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSize
StSize
StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7
8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step
2
Step
3
MOD
2nd
Help
RF
ON/OFF
NSG 4070
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp in
Amp out
50 Ω Termination
NSG 4070
RF output
NSG 4070
User port
Amp in
Amp out
Power amplifier
inputs
Calibration jig
BCI probe
Ground plane
Interlock
Switch (part of SW 4070)
Switch (part of SW 4070)
2
A
A
1
B
B
DIRECTIONAL COUPLER
Attenuator #1 Attenuator #2
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSize
StSize
StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7
8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step
2
Step
3
MOD
2nd
Help
RF
ON/OFF
NSG 4070
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp in
Amp out
NSG 4070
RF output
NSG 4070
User port
Amp in
Amp out
Power amplifier
inputs
BCI probe
Ground plane
Interlock
Switch (part of SW 4070)
Switch (part of SW 4070)
2
A
A
1
B
B
DIRECTIONAL COUPLER
Attenuator #1 Attenuator #3
Auxiliary
equipment
AE
EUT
Equipment
under test
Insulating
MD 4070
Test level
for MIL STD 461F CS114 curve #3
Calibration set-up (for curve #3)
Test set-up with monitoring probe (for curve #3)
Attenuator #1: 10 dB, 30 W
Attenuator #2: not in use
Attenuator #3: not in use
Termination: 50 Ω 10 W
Remarks:
Calibration with a 10 dB increased
level.
Attenuator #1: 10 dB, 30 W
Attenuator #2: not in use
Attenuator #3: 10 dB, 30 W
Termination: 50 Ω 10 W
Remarks:
Test with additional 10 dB attenu-
ator (attenuator #3) between
directional coupler and BCI probe.
Use of MD 4070 on PM ch.1:
The monitoring probe is in the
linear measuring range above 57
dBμA stress level (above 30 kHz).
The MD 4070 needs to be used in
the active mode.
49
89
81
89
30
40
50
60
70
80
90
100
110
0,01
1
30
200
Frequency in MHz
Current in dBuA