beautypg.com

Ds 5630 dips and drops switch – Atec Teseq-Schaffner-NSG-5600 User Manual

Page 6

background image

The DS 5630 switches the voltage source through to
the EUT connection under program-control. The
primary input and the auxiliary connection enable
two sources to be used. The DC switching capacity
of 70V/75A is capable of coping with high power
requirements and is ready for the 42V era.

The selectable switching conditions are:
Output (EUT)

Q

to primary source

Q

to auxiliary source

Q

switched off (open)

Q

dip and drop from primary to auxiliary source
and back again

Q

to primary source with 2

Ω extra impedance

(SAE J 1113-11 pulse 2b)

Due to the rapid switching times, the requirements
of every standard that calls for drop-outs and volt-
age jumps in the µs range can be fulfilled by using
a battery simulator and an auxiliary source. The
DS 5630 also acts as the DC input for the transformer
coupled sine wave noise or for magnetic field im-
munity tests when high field strengths are required.

DS 5630
Dips and drops switch

A powerful electronic switch for fast dips and drops applications.