Switching actions with inductive and other loads
influenced by complex inductances of the wiring
harness all create disturbances that must be
simulated. ISO and SAE have defined these tests
as pulse 1, 2 and 6. The MT 5510 produces these
test pulses in conformance with the relevant

standard. In order to be able to cope with a com-
pany’s own test methods, the generator module also
provides a much greater range of selectable para-
meters, considerably higher test voltages, additional
impedances and pulse widths.

Pulse

Amplitude

Impedance

Rise time

Pulse duration

Repetition

[V]

[

Ω

Ω]

[µs]

[s]

Pulse 1

-20 to -200

10, 20, 50

1 or 3

2ms

0.5 - 5

-150 to -600

20

1

1ms

0.5 - 5

-12 to -220

2, 10, 50

1

50µs

0.2 - 5

Pulse 1 (24V)

-150 to -600

50

3

1ms

0.5 - 5

Pulse 2

12 to 220

2, 10, 20, 50

1

50µs

0.2 - 5

Pulse 6

-10 to -350

30

<60

300µs

0.5 - 15

CI 220

-20 to -110

10

1

2ms

0.5 - 5

-30 to -300

4

1

50µs

0.2 - 5

20 to 150

4

1

50µs

0.2 - 5

Technical specifications MT 5510

MT 5510
Transient generator
ISO pulse 1, 2, 6 and variants