PCU-LT/PCU1-SP Control Unit Specification

Loading Unit Current Metering

The AC output current is metered by a true RMS memory

ammeter (acquisition time 200ms) with a liquid crystal display.

The current metering has 3 ranges corresponding to 10%, 50%

and 100% of the maximum rating of the loading unit. In addition,

a 200% metering range is enabled in pulse mode.

NLU2000

Range Full scale Resolution Accuracy

10% 200.0A 0.1A ±0.5%rdg+5d*
50% 1000A 1A ±0.5%rdg+5d*
100% 2000A 1A ±0.5%rdg+5d*
200% 4000A 1A ±1%rdg+5d**

NLU5000

Range Full scale Resolution Accuracy

10% 500.0A 0.1A ±0.5%rdg+5d

*1

50% 2500A 1A ±0.5%rdg+5d

*1

100% 5000A 1A ±0.5%rdg+5d

*1

200% 10kA 10A ±1.5%rdg+5d

*2

*1

1.5% rdg+5d in pulse mode

*2

selected in pulse mode

Timing System

The PCU1 systems have a flexible timing system with two

contact inputs and 5 operating modes. Each contact circuit

automatically selects for N/O or N/C contacts, and the status of

each contact input is shown by an LED. The timing channels

may also be triggered by a dc voltage between 24 and 240V.

Timer resolution

1ms

Timer full scale

999.999s

Timer accuracy

±0.01%rdg+2d (4d current mode)

Contact O/C voltage

24V

Contact S/C current

20mA

Vdc input range

24-240Vdc

Timer mode Timer start Timer stop

Internal Start ‘On’ button Contact

Single contact Contact 1 Contact 1

Dual contact Contact 1 Contact 2

Current operated

*3

Current >20% rng Current <20% rng

Pulse mode 0.2s

*4

‘On button’ 0.2s

Pulse mode 0.5s

*4

‘On button’ 0.5s

Pulse mode 1s

*4

‘On button’ 1s

Pulse mode 2s

*4

‘On button’ 2s

Off Setting position

*3

Current operated mode

is used to time circuit breakers with no

auxiliary contacts. The timer is started when the current exceeds

20% of the selected metering range (e.g. 100A on the NLU5000

500A range). The timer stops when the current falls.

*4

Pulse mode

applies current to the load for a maximum of the

specified time. If contact set 1 changes state or the current

drops below 20% of the metering range during the pulse ti me,

the timer is stopped. The maximum output current is increased

in pulse mode. The maximum obtainable current is set by the

impedance of the test object and output leads.

Secondary Injection Output (PCU1-SP only)

The secondary current injection output on the unit has two taps,

allowing the injection of currents up to 100A.

Output Range Continuous Intermittent current
current 5min on

*5

1 min on

*5

0-5V 33A 67A 100A
0-16V 10A 20A 30A

*5

All on times must be followed by an off time of 15 minutes

Metering Range Resolution Accuracy Current trip

10.00A 0.01A ±0.5%rdg

+5d 10.5A

20.00A 0.01A ±0.5%rdg

+5d 21A

100.0A 0.1A ±0.5%rdg

+5d 100A

The output is protected by electronic trips and a circuit breaker.

Supply Requirements

230V±10%, 45-65Hz 1ph 11.5kVA max (23kVA overload for 2s)

Control Unit Standard Accessories

Mains lead (5m), loading unit power and metering leads (5m),

operating manual and spare fuses.

Dimensions

PCU1-SP/PCU1-LT 450(w) x 275(h) x 305mm(d)

NLU2000/NLU5000 450(w) x 275(h) x 370mm(d) incl. terminals

Weight

PCU1-LT 23kg PCU1-SP 26kg

NLU5000 55kg NLU2000 50kg

Temperature Range

Storage -20°C to 60°C, Operating 0°C to 45°C

Protection and Safety

The PCU1 series and loading units are CE marked and are

designed to meet the requirements of BS EN61010. The system

is protected by electronic trips on the outputs, circuit breakers on

the mains input, control unit output and secondary injection

output (PCU1-SP only). The unit also has a duty cycle trip on

the loading unit output and the loading unit has comprehensive

thermal protection.

Optional Loading Unit Specifications

Two loading units are available to provide a range of output

currents suitable for different primary injection tasks. Each

loading unit has three output taps allowing current injection into a

wide range of loads of differing impedances. Optional output

lead sets are also available (see www.trtest.com for details).

NLU5000 Loading Unit Intermittent Ratings

Output Maximum current
Voltage* Cont. 5 min 1 min 40s

2.3V

1500A 3000A 4500A 5000A

4.6V

750A 1500A 2250A 2500A

9.2V

375A 750A 1125A 1250A

NLU2000 Loading Unit Intermittent Ratings

Output Maximum current
Voltage* Cont. 5 min 1 min 40s

3.5V

600A 1200A 1800A 2000A

6.9V

300A 600A 900A 1200A

13.8V

150A 300A 450A 500A

I<