Dc voltage measurement, Ac voltage measurement, Dc current measurement – Atec Fluke-740 Series User Manual
Page 8: Resistance measurement, Frequency measurement, Dc voltage output, Dc current output, Resistance sourcing, Frequency sourcing
![background image](/manuals/319614/8/background.png)
DC voltage measurement
Accuracy (% of reading + % of full scale)
Range (full scale)
1 year
2 years
110.000 mV
0.025% + 0.015%
0.05% + 0.015%
1.10000 V
0.025% + 0.005%
0.05% + 0.005%
11.0000 V
0.025% + 0.005%
0.05% + 0.005%
110.000 V
0.05% + 0.005%
0.1% + 0.005%
300.00 V
0.05% + 0.005%
0.1% + 0.005%
Temperature coefficient: (0.001% reading + 0.0015% f.s.)/°C
from -10 °C to 18 °C and 28 °C to 50 °C
Input impedance: 5 MΩ
Common mode error: 0.008% f.s./(Common Mode Volt)
Maximum input voltage: 300 V rms
AC voltage measurement
Accuracy (% of reading + counts)
Frequency range
1 year
2 years
20 to 40 Hz
2% + 10
2% + 10
40 to 500 Hz
0.5% + 5
0.5% + 5
500 Hz to 1 kHz
2% + 10
2% + 10
1 to 5 kHz
10% + 20
10% + 20
Ranges: 1.1000, 11.000, 110.00, 300.0 V
Specifications apply for 10% to 100% of range
Input impedance: 5 MΩ and <100 pF
Input coupling: AC only
Temperature coefficient:
10% of spec/°C from -10 °C to 18 °C and 28 °C to 50 °C
Maximum input voltage: 300 V rms
Minimum input voltage: 0.5 V above 1 kHz
DC current measurement
Accuracy (% of reading + % of full scale)
Range (full scale)
1 year
2 years
30.000 mA
0.01% + 0.015%
0.02% + 0.015%
110.00 mA
0.01% + 0.015%
0.02% + 0.015%
Temperature coefficient: (0.001% reading + 0.002% f.s.)/°C
from -10 °C to 18 °C and 28 °C to 50 °C
Common mode error: 0.01% f.s./(Common Mode Volt)
Maximum input voltage: 30 V dc
Resistance measurement
Accuracy (% of reading + ohms)
Range (full scale)
1 year
2 years
11.000 Ω
0.05% + 50 mΩ
0.075% + 50 mΩ
110.00 Ω
0.05% + 50 mΩ
0.075% + 50 mΩ
1.1000 kΩ
0.05% + 0.5 Ω
0.075% + 0.5 Ω
11.000 kΩ
0.1% + 10 Ω
0.1% + 10 Ω
Temperature coefficient: (0.01% f.s. + 2 mΩ)/°C from -10 °C to 18 °C
and 28 °C to 50 °C
Common mode error: 0.005% f.s./(Common Mode Volt)
Maximum input voltage: 30 V dc
Continuity: Continuous tone <25 Ω, No tone >400 Ω
Frequency measurement
Accuracy
Range
1 year
2 years
1.00 to 109.99 Hz
0.05 Hz
0.05 Hz
110.0 to 1099.9 Hz
0.5 Hz
0.5 Hz
1.100 to 10.999 kHz
0.005 kHz
0.005 kHz
11.00 to 50.00 kHz
0.05 kHz
0.05 kHz
For frequencies <109.99 Hz, specification applies for signals with
slew rates >5 V/ms
Minimum amplitude for Hz measurement: (Squarewaves) 1 Hz to
1 kHz, 300 mV p-p; 1 kHz to 30 kHz, 1.4 V p-p; >30 kHz, 2.8 V p-p
Maximum input: 1 Hz to 1 kHz, 300 V rms; >1 kHz, 30 V rms
Input impedance: 5 MΩ
Measurement Function Specifications
Sourcing (Simulation) Function Specifications
DC voltage output
Accuracy (% of output + % of full scale)
Range (full scale)
1 year
2 years
110.000 mV
0.01% + 0.005%
0.015% + 0.005%
1.10000 V
0.01% + 0.005%
0.015% + 0.005%
15.0000 V
0.01% + 0.005%
0.015% + 0.005%
Temperature coefficient: (0.001% output + 0.001% f.s.)/°C
from -10 °C to 18 °C and 28 °C to 50 °C
Maximum output current: 10 mA
Loading: (0.001% f.s. +1 µV)/mA
Common mode error: 0.008% f.s./(Common Mode Volt)
Maximum input voltage: 30 V dc
DC current output
Accuracy (% of output + % of full scale)
Range (full scale)
1 year
2 years
22.000 mA
0.01% + 0.015%
0.02% + 0.015%
Current sink (sim-
0.02% + 0.03%
0.02% + 0.03%
ulate transmitter)
Specification applies from 2 to 22 mA; below 2 mA typical accuracy is
0.15% of full scale
Maximum burden voltage: 24 V
Temperature coefficient: (0.003% output + 0.003% f.s.)/°C
from -10 °C to 18 °C and 28 °C to 50 °C
Common mode error: 0.008% f.s./(Common Mode Volt)
Maximum input voltage: 30 V dc
Resistance sourcing
Accuracy (% of output + ohms)
Range (full scale)
1 year
2 years
11.000 Ω
0.01% + 20 mΩ
0.02% + 20 mΩ
110.00 Ω
0.01% + 40 mΩ
0.02% + 40 mΩ
1.1000 kΩ
0.02% + 0.5 Ω
0.03% + 0.5 Ω
11.000 kΩ
0.03% + 5 Ω
0.04% + 5 Ω
Temperature coefficient: 0.01% f.s./°C from -10 °C to 18 °C and
28 °C to 50 °C
Maximum and minimum current through source resistance:
Maximum
Minimum
11 Ω range:
3 mA dc
0.1 mA dc
110 Ω range:
3 mA dc
0.1 mA dc
1.1 kΩ range:
3 mA dc
0.01 mA dc
11 kΩ range:
1 mA dc
0.01 mA dc
Common mode error: 0.008% f.s./(Common Mode Volt)
Maximum input voltage: 30 V dc
Frequency sourcing
Accuracy
Range
1 year
2 years
0.00 to 10.99 Hz
0.01 Hz
0.01 Hz
11.00 to 109.99 Hz
0.1 Hz
0.1 Hz
110.0 to 1099.9 Hz
0.1 Hz
0.1 Hz
1.100 to 21.999 kHz
0.002 kHz
0.002 kHz
22.000 to 50.000 kHz
0.005 kHz
0.005 kHz
Waveforms: Squarewave w/50% duty cycle, sinewave
Amplitude: 0.1 to 10 V p-p
Amplitude accuracy: 3% of output + 0.5% of f.s., 1 to 1099 Hz; 10%
of output + 0.5% of f.s., 1.1 to 10.9 kHz; 30% of output + 0.5% f.s.,
11 to 50 kHz
Maximum input voltage: 30 V dc
8
Fluke Corporation
740 Series Brochure