6 5502A Multi-Product Calibrator

Function and range
Direct volts

0 to ± 1020 V

Direct current

0 to ± 20.5 A

Alternating volts

1 mV to 1020 V

10 Hz to 500 kHz

Volt/hertz

1000 V@ 10 kHz/330 V@100 kHz

Alternating current

29 µA to 20.5 A

10 Hz to 30 kHz

Waveforms

Sine, square, triangle, truncated sine

Resistance

0 Ω to 1100 MΩ

Capacitance

220 pF to 110 mF

Power (phantom loads)

20.9 kW

Phase control

0.01°

Thermocouple (source and

measure temperature)

B, C, E, J, K L N R, S, T, U

10 µV/°C and 1 mV/°C

RTD (source temperature)

Pt 385-100 Ω, Pt 3926-100 Ω Pt 3916-100 Ω, Pt 385-200 Ω,

Pt 385-500 Ω, Pt 385 1000 Ω, PtNi 385-120 Ω (Ni120),

Cu 427 10 Ω

Interfaces

RS-232, IEEE 488

Frequency uncertainty

< 25 ppm

Oscilloscope calibrator (options)

Levelled sine wave from 5 mV to 5.5 Vpp max, frequencies

50 kHz to 600 kHz; edge rise times of < 300 ps, multiple trig-

ger functions, lowest dc, square wave and timing uncertainty

Amplified current

(accessory amplifier)

Extend from 20.5 A to a maximum of 100 A dc and 120 A ac

from 10 Hz to 10 k Hz

Summary

specifications

Calibrate

almost

anywhere

Rugged transit

case makes on-site

calibration safe,

easy, convenient

An optional shock-mounted

transit case featuring built-

in handles and wheels gives

you the option of taking the

calibrator to the workload

for on-site or mobile appli-

cations. Once at the site,

just remove the front and

rear doors from the case

for access to the 5502A’s

front and rear panels—the

top, bottom and sides of the

calibrator remain protected,

and you don’t need to fully

unpack the calibrator for

each use.