Pressure specifications, Pressure module specifications, The fluke family of 29 pressure modules – Atec Fluke-750 Series User Manual
Page 10
Reference
High
2
Low
2
Max over-
Range/
Range (approx)/ uncertainty Stability Temperature
Total
1
side
side
Fitting
pressure
Model
Resolution Resolution (23 ± 3 ºC)
(1 year)
(0 to 50 ºC) uncertainty media media material (x nominal)
Differential
FLUKE-700P00 1 in. H
2
O/0.001 0.25 kPa/0.0002
0.300
0.025
0.025
0.350
Dry
Dry
316 SS
30x
FLUKE-700P01 10 in. H
2
O/0.01
2.5 kPa/0.002
0.200
0.050
0.050
0.300
Dry
Dry
316 SS
3x
FLUKE-700P02
1 psi/0.0001
6900 Pa/0.7
0.150
0.070
0.080
0.300
Dry
Dry
316 SS
3x
FLUKE-700P22
1 psi/0.0001
6900 Pa/0.7
0.100
0.020
0.030
0.150
316 SS
Dry
316 SS
3x
FLUKE-700P03
5 psi/0.0001
34 kPa/0.001
0.050
0.020
0.030
0.100
Dry
Dry
316 SS
3x
FLUKE-700P23
5 psi/0.0001
34 kPa/0.001
0.025
0.010
0.015
0.050
316 SS
Dry
316 SS
3x
FLUKE-700P04
15 psi/0.001
103 kPa/0.01
0.025
0.010
0.015
0.050
Dry
Dry
316 SS
3x
FLUKE-700P24
15 psi/0.001
103 kPa/0.01
0.025
0.010
0.015
0.050
316 SS
Dry
316 SS
3x
Gage
FLUKE-700P05
30 psi/0.001
207 kPa/0.01
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700P06
100 psi/0.01
690 kPa/0.07
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700P27 300 psi / 0.01
2070 kPa / 0.1
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700P07
500 psi/0.01
3400 kPa/0.1
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700P08
1000 psi/0.1
6900 kPa/0.7
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700P09
1500 psi/0.1
10 MPa/0.001
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
2x
Absolute (not compatible with Fluke 701 or 702)
FLUKE-700PA3
5 psi/0.0001
34 kPa/0.001
0.050
0.010
0.010
0.070
316 SS
N/A
316 SS
3x
FLUKE-700PA4
15 psi/0.001
103 kPa/0.01
0.050
0.010
0.010
0.070
316 SS
N/A
316 SS
3x
FLUKE-700PA5
30 psi/0.001
207 kPa/0.01
0.050
0.010
0.010
0.070
316 SS
N/A
316 SS
3x
FLUKE-700PA6
100 psi/0.01
690 kPa/0.07
0.050
0.010
0.010
0.070
316 SS
N/A
316 SS
3x
Vacuum (not compatible with Fluke 701 or 702)
FLUKE-700PV3 -5 psi/0.0001
-34 kPa/0.001
0.040
0.015
0.015
0.070
316 SS
Dry
316 SS
3x
FLUKE-700PV4 -15 psi/0.001
-103 kPa/0.01
0.040
0.015
0.015
0.070
316 SS
Dry
316 SS
3x
Dual
FLUKE-700PD2 ±1 psi/0.0001
±6900 Pa/0.7
0.150
0.025
0.025
0.200
316 SS
Dry
316 SS
3x
FLUKE-700PD3 ±5 psi/0.0001
±34 kPa/0.001
0.040
0.015
0.015
0.070
316 SS
Dry
316 SS
3x
FLUKE-700PD4 ±15 psi/0.001
±103 kPa/0.01
0.025
0.010
0.015
0.050
316 SS
Dry
316 SS
3x
FLUKE-700PD5 -15/30 psi/0.001 -100/207 kPa/0.01
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700PD6 -15/100 psi/0.01 -100/690 kPa/0.07
0.025
0.010
0.015
0.050
316 SS
N/A
316 SS
3x
FLUKE-700PD7 -15/200 psi/0.01 -100/1380 kPa/0.1
0.040
0.015
0.015
0.070
316 SS
N/A
316 SS
3x
High
FLUKE-700P29
3000 psi/0.1
20.7 M Pa/0.001
0.050
0.010
0.020
0.080
C276
N/A
C276
2x
FLUKE-700P30
5000 psi/0.1
34 M Pa/0.001
0.050
0.010
0.020
0.080
C276
N/A
C276
2x
FLUKE-700P31
10000 psi/1
69 M Pa/0.007
0.050
0.010
0.020
0.080
C276
N/A
C276
1.5x
Pressure specifications
The Fluke family of 29 pressure modules:
Covers virtually any pressure application including
gage, differential, dual (compound), absolute, and
vacuum.
• Display pressure readings in any of ten different
pressure units you specify in the calibrator setup.
• Rugged urethane molded cases protect the modules
from rough handling and harsh conditions.
• Features internal temperature compensation from
O °C to 50 °C for full-accuracy performance.
• Includes NIST-traceable calibration certificate.
• Modules can be calibrated locally, helping to
control costs.
Pressure module specifications
(all specifications in % of full span. Specifications reflect a confidence interval of 95 %.)
1
Total uncertainty, one year for temperature range 0 °C to +50 °C. Total uncertainty, 1.0% of full span for temperature range -10 °C to 0 °C. For P00 module only, compensated
temperature range is 15 ° to 35 °C.
2
“Dry” indicates dry air or non-corrosive gas as compatible media. “316 SS” indicates media compatible with Type 316 Stainless Steel. “C276” indicates media compatible with
Hastelloy C276.
Use of pressure zero is required prior to measurement or source. Maximum overpressure specification includes common mode pressure. Modules are rated. Metric adapter(s):