beautypg.com

Specifications – Atec Fluke-9190A User Manual

Page 3

background image

3 Fluke Calibration 9190A Ultra-Cool Field Metrology Well

Specifications

Base unit specifications
Temperature range at 23 °C

–95 °C to 140 °C (–139 °F to 284 °F)

Display accuracy

± 0.2 °C full range

Accuracy with external reference

[3]

± 0.05 °C full range

Stability

± 0.015 °C full range

Axial uniformity at 40 mm (1.6 in)

± 0.05 °C full range

Radial gradient

± 0.01 °C full range

Loading effect

(with a 6.35 mm reference probe and three 6.35 mm probes)
± 0.006 °C full range
(versus display with one 6.35 mm probe)
± 0.25 °C at –95 °C
± 0.10 °C at 140 °C

Operating conditions

0 °C to 35 °C, 0 % to 90 %
RH (non-condensing) < 2000 m altitude

Environmental conditions for all

specifications except temperature range

13 °C to 33 °C

Immersion (well) depth

160 mm (6.3 in)

Well diameter

30 mm (1.18 in)

Heating time

[1]

–95 °C to 140 °C: 40 min

Cooling time

[1]

23 °C to –90 °C: 80 min
23 °C to –95 °C: 90 min
140 °C to 23 °C: 60 min

Stabilization time

[2]

15 min

Resolution

0.01

Display

LCD, °C or °F user selectable

Size (H x W x D)

480 mm x 205 mm x 380 mm (18.8 in x 8.0 in x 14.9 in)

Weight

16 kg (35 lb)

Power requirements

100 V to 115 V (± 10 %) 50/60 Hz, 575 W
200 V to 230 V (± 10 %) 50/60 Hz, 575 W

System fuse ratings

115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V

4–20 mA Fuse (-P model only)

50 mA F 250 V

Computer interface

RS-232, USB Serial, and 9930 Interface-It temperature

calibration software included

Safety

IEC 61010-1, Installation Category II, Pollution degree 2

Electromagnetic environment

IEC 61326-1: Basic

Refrigerants

R32 (Difluoromethane)
< 20 g, ASHRAE safety group A2L
R704 (Helium)
< 20 g, ASHRAE safety group A1