S310 specifications – Atec Sencore-S310 User Manual
Page 2
S310 Specifications
--------------------------------------------------------------------------------
MEASUREMENT SPECIFICATIONS
Serial Signals
Amplitude - INPUT SIGNAL MAXIMUM LIMITS: ±10 V. MEASUREMENT RANGE: ±1.5 V. RIPPLE
MEASUREMENT RANGE (Vrp): <1.5 V. Vrp FREQUENCY RANGE: 0 to 10 kHz. MEASUREMENT
RESOLUTION: 12 mV. TEST OUTPUT FOR MEASUREMENTS: Vpp and Vrp identical value to voltage
measured at 75 Ohms.
Jitter - PARALLEL SIGNAL JITTER RANGE: ±7.5 ns. SERIAL SIGNAL JITTER RANGE: ±1.5 ns. SE-
RIAL SIGNAL JITTER RESOLUTION: 40 pS +10 pS with 10 m of cable. PARALLEL SIGNAL JITTER
BANDWIDTH: 1.4 kHz. PARALLEL TEST OUTPUT JITTER: 100 mV/ns at 75 Ohms. SERIAL TEST OUT-
PUT JITTER: 300 mV/ns at 75 Ohms. MISCELLANEOUS - MON OUTPUT: 720 mV at 75 Ohms, DC com-
ponent not transmitted. CLOCK OUTPUT: Parallel signal clock 400 mVpp at 75 Ohms, DC component not
transmitted.
Measurements Common To Serial And Parallel Signals - PREAMBLE ERROR DETECTION: The analyser
has a “hard” preamble detector, which allows detection of a preamble error on the 4 least significant bits. TEST
OUTPUT LEVELS FOR DISPLAY OF RESERVED CODES AND TRS ERRORS: 700 mV at 75 Ohms.
Major Alarm Threshold
SERIAL SIGNAL: Signal absence or TRS word error frequency of more than 10 errors in 500 ms. PARALLEL
SIGNAL: When clock/data phase measurement shows clock phase to be outside the range of ±11 nS. SECOND-
ARY ALARM THRESHOLD: When one parameter falls outside specifications. SERIAL LINK: Adjustable
from 2,400 to 19,200 Baud; delivered set at 9,600 Baud. OTHER PARAMETERS: 8 bits, 1 start bit, no parity.
Parallel Signals
Amplitude - INPUT SIGNAL MAXIMUM LIMITS: 0 V to -5.2 V. CLOCK SIGNAL MEASUREMENT
RANGE: -1.29 V ± 1.29 V. VAV DATA MEASUREMENT RANGE: -1.29 V ±1.29 V. MEASUREMENT
RESOLUTION: 11mV. MEASUREMENT PRECISION: ±25 mV. TEST OUTPUT FOR VPP AND VRP MEA-
SUREMENTS: Vpp value identical to the measured signal. VAV: Displayed value is measured value +1.79V.
Clock/Data Phase - PRECISION: ±1.5 ns. TEST OUTPUT LEVEL: 700 mV at 75 Ohms.