beautypg.com

Atec HP-Agilent-35665A User Manual

Page 4

background image

Software Limits at 5 < or = RPM < or = 491,519;
and Number of Orders < = 200
Delta Order: 1/128 to 1/1

Resolution
(Maximum Order) / (Delta Order): <200

RPM Accuracy: ±100 ppm

Maximum RPM Ramp Rate: 750 RPM/second (typical)
1 and 2 Channel Mode
1000 to 10,000 RPM Run-up
Maximum Order = 10
Delta Order = 0.1
RPM Step = 30

REAL TIME OCTAVE ANALYSIS - OPTION 1D1

STANDARDS

ANSI Standard S1.11 - 1996, Order 3, Type 1-D, Extended and
Optional Frequency Ranges
IEC Standard 651-1979 Type 0 Impulse, and ANSI S1.4

1/1 OCTAVE RANGE (at centers): 0.0613 Hz to 16 kHz

1/3 OCTAVE RANGE (at centers): 0.08 Hz to 32 kHz

1/12 OCTAVE RANGE (at centers): 0.09145 Hz to 12.338 kHz

1/3 OCTAVE UPDATE RATE
1 Channel Mode
16 kHz Bandwidth 12 Updates/Second (Typical)
32 kHz Bandwidth 5 Updates/Second (Typical)
2 Channel Mode 8 Updates/Second (Typical)
(16 kHz, 32 kHz Bandwidths)

1/3 OCTAVE DYNAMIC RANGE
2 Second Stable Average: >80dB as per ANSI S1.11-1986
Totalpower limited by the for full scale ranges from
input noise level 50 m Vpeak to 31.7 Vpeak

STABLE AVERAGE RANGE
One Channel Mode Real-Time Bandwidth = 32 kHz
Two Channel Mode Real-Time Bandwidth = 16 kHz

SWEPT SINE MEASUREMENTS - OPTION 1D2

DYNAMIC RANGE: >130 dB (typical)
Input auto range on
100 msec integration

SWEEP TIMES:
51.2 Hz to 51.2 kHz span
11 dBV rms source level