2400 series technical specifications, Modulation specifications: m/am series only – Atec Giga-tronics-2420L-2420M User Manual
Page 3
2400 Typical Frequency Settling Time
200
400
0
0.001
0.01
0.1
1.0
4.0
600
Time in
µ
se
c
2400 S ries SSB Phase Noise
-150
-140
-130
-120
-110
-50
10
100Hz
1kHz
10kHz
100kHz
1MHz
Hz
10MH
10 GHz
-100
-60
-70
-80
-90
-160
L(f) dBc/Hz
-40
Offset frequency
2400 Series
Technical Specifications
All specifications apply over a 0˚C to +55˚C range after 30 minutes of warm-up time unless otherwise stated.
SSB Phase Noise
Frequency
Offset from Carrier (dBc/Hz)
(GHz) 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz
0.85
1.85
5.6
10
18
23
30
-92
-86
-75
-74
-68
-63
-61
-111
-105
-97
-92
-89
-85
-83
-112
-106
-98
-92
-90
-86
-84
-123
-117
-105
-101
-99
-93
-91
-130
-135
-130
-128
-123
-118
-115
Frequency
(GHz)
50 Hz - 15 kHz
Bandwidth
.01 – 16 GHz
16 – 32 GHz
32 – 40 GHz
< 40 Hz
< 80 Hz
< 120 Hz
Frequency
(GHz)
Offsets
> 5 MHz
.01 – 2 GHz
2 – 20 GHz
20 – 40 GHz
– 130 dBm/Hz
– 145 dBm/Hz
– 140 dBm/Hz
Spectral Purity Continued:
Residual FM (typical)
AM Noise (typical)
Frequency & Power Sweep - L/M Series only (Option 43)
Ramp Frequency Sweep:
Ramp Power Sweep:
Power Slope:
Power Flatness:
Ramp Output:
Z-Axis Blanking:
Sweep Time
4
:
Full Frequency Coverage
0 to 25 dB
0 to 0.5 dB/GHz
See Accuracy table
0 to 10V
± 5V
10 msec — 30 secs
List Mode
Number of List Points:
4000
Frequency Settling Time
5
:
< 550 µsec for ∆F0
6
≤ 500 MHz
Amplitude Settling Time
7
:
< 500 µsec
Step Time:
150 µsec - 1 sec
2 ms - 1 sec (option 31)
Sync Out Delay
8
:
50 µsec - 10 msec
Trigger Modes:
EXT, GPIB GET, Software
List Mode Continued:
Sweep Modes:
Continuous, Single Step,& Single Sweep
Remote Programming
Hardware Interface:
IEEE 488.2
RS-232 & USB (w/supplied adapter)
Software Interface:
SCPI & GT-12000A
Automation Xpress Interface (option 48)
Execution Speed (IEEE 488.2):
CW Switching
4000 pt.
download
2.5 ms
20 sec
28 ms
28 sec
22 ms
288 sec
AXI
SCPI
GT-12000A
Amplitude Modulation
9
Depth
10
:
0 to 75% (Level = 0 dBm)
Rate
(3 dB Bandwidth)
:
DC — 5 kHz (depth = 30%)
Sensitivity:
0 — 95% /V selectable
Accuracy:
± 10% of setting at 1 kHz rate, 0 - 90%/V
Input:
Range:
± 1 V
Impedance:
600Ω
Scan Modulation
(Specification applies for frequencies below 20 GHz)
Depth:
> 60 dB
Scan Time:
200 ms - 10 sec
Maximum Number of Points: 4000
Minimum Time per Point:
1 ms
Scan Pattern:
Sin (x)
X
Minimum # of Lobes:
1
Modulation Specifications: M/AM Series Only
∆F0 in GHz
6
Automation Xpress Interface (AXI) (Option 48)
For use with Giga-tronics Automation Xpress software. The AXI
provides Xpress 2.5 ms CW Frequency/Power switching, faster data
exchange and functional downloads/executions, and a stable API
programming interface for the ATE programming environment.
3
4 Sweep rate must be < 500 MHz/msec.
5 Time for frequency to settle within 50 kHz of final value after a frequency switch.
6
∆
F0 = | (Fstop x Nstop) – (Fstart x Nstart) | - See Frequency Bands Table for N values.
7 Time for amplitude to settle within 0.1 dB of final value after an amplitude switch.
8 Delay is specified from edge of trigger pulse.
9 Modulation peaks must be less than maximum available power.
10 Levels noted can be offset using step attenuator (option 26).