Lock-in Amplifiers

Model 7280 Specifications

Outputs
Main Analog (CH1 and CH2) Outputs

Function

X, Y, R,

θ, Noise,

Ratio, Log Ratio and
User Equations 1 & 2.

Amplitude

±2.5 V full-scale;
linear to ±300% full-
scale

Impedance

1 k

Update Rate:
X, Y or R @ TC

≤4 ms 7.5 MHz

All outputs @ TC

≥5 ms 1 kHz

Signal Monitor

Amplitude

±1 V FS

Impedance

1 k

Auxiliary D/A Output 1 and 2

Maximum Output

±10 V

Resolution

1 mV

Accuracy

±10 mV

Output Impedance

1 k

8-bit Digital Port

0 to 8 lines can be configured as inputs,
with the remainder being outputs. Each
output line can be set high or low and
each input line read to allow interaction
with external equipment. Extra line acts
as trigger input

Reference Output

Waveform

0 to 3 V rectangular
wave

Impedance

TTL-compatible

Power - Low Voltage

±15 V at 100 mA rear
panel 5-pin 180° DIN
connector for
powering
SIGNAL RECOVERY
preamplifiers

Data Storage Buffer

Size

32k × 16-bit data
points, may be
organized as 1×32k,
2×16k, 3×10.6k, 4×8k,
etc.

Max Storage Rate
From LIA

up to 1000 16-bit
values per second

From ADC1

up to 40,000 16-bit
values per second

User Settings

Up to 8 complete instrument settings
can be saved or recalled at will from
non-volatile memory

LabVIEW Driver Software

A LabVIEW driver for the instrument is
available from the
www.signalrecovery.com website,
offering example VIs for all its controls and
outputs, as well as the usual Getting
Started and Utility VIs. It also includes
example soft-front panels built using these
VIs, demonstrating how you can
incorporate them in more complex
LabVIEW programs.

SIGNAL RECOVERY Acquire
Software (see page 56)

Users who do not wish to write their own
control code but who still want to record
the instrument’s outputs to a computer file
will find the SIGNAL RECOVERY Acquire
Lock-in Amplifier Applications Software,
available at a small extra cost, useful. This
32-bit package, suitable for Windows XP/
Vista, extends the capabilities of the
instrument by, for example, adding the
ability to record swept frequency
measurements. It also supports the
internal curve buffer, allowing acquisition
rates of up to 1000 points per second
independent of the computer's processor
speed.

Interfaces

RS232 and GPIB (IEEE-488). A second
RS232 port is provided to allow “daisy-
chain” connection and control of up to 16
units from a single RS232 computer port

General
Power Requirements

Voltage

110/120/220/240 VAC

Frequency

50/60 Hz

Power

200 VA max

Dimensions

Width

17¼" (435 mm)

Depth

19" (485 mm)

Height
With feet

6" (150 mm)

Without feet

5¼" (130mm)

Weight

25.4 lb (11.5 kg)

Ordering Information

Each model 7280 is supplied complete with a comprehensive instruction manual. Users
may download the instrument's LabVIEW driver software and a free demonstration
copy, DemoAcquire, of the SIGNAL RECOVERY lock-in amplifier applications software
package, from the www.signalrecovery.com website.

Optional Accessories

Model 7280/99

Extended frequency range (800 kHz) for Dual Reference and
Dual Harmonic Modes

Model 7280/98

Extended frequency range (2.0 MHz) for Dual Reference and
Dual Harmonic Modes

Acquire™

32-bit lock-in amplifier applications software for use with
Windows XP/Vista operating systems

Model K02004

Rack mount to mount one model 7280 in a 19" rack