Specifications – Atec NHResearch-4700Series User Manual
Page 4
1
Specifications apply at 23
o
+/- 5
o
C after a 10 minute warm up.
2
Accuracies apply when Settings and/or Measurements >10% of Range.
3
Current linearly reduced between 1 and 0.15 V.
4
Models 2 - 36 kW also have a 20 A /1 KW Range with reduced accuracy.
5
Set 1000% to 6000% of Range = 10% Accuracy.
6
Single channel capture. Simultaneous Voltage and Current captures would halve sample
rate and memory available.
SPECIFICATIONS
1
4700 Ratings
4700-1
4700-2
4700-3
4700-6
4700-12
4700-18
4700-24
4700-36
Power
1 kW
2 kW
3 kW
6 kW
12 kW
18 kW
24 kW
36 kW
Maximum Current
2
200 A
400 A
600 A
1200 A
2400 A
3600 A
4800 A
7200 A
Voltage Range
3
1-120 V
1-120 V
1-120 V
1-120 V
1-120 V
1-120 V
1-120 V
1-120 V
Programmable Modes
Accuracies: % of Set + % of Range, Resolution: % of Range
Constant Current
Ranges
4
20, 200 A
40, 400 A
60, 600 A
120, 1200 A
240, 1200 A
360, 3600A
480, 4800 A
720, 7200 A
Accuracy
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
0.12%+0.08%
Resolution
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Constant Voltage
Ranges
6.6, 20, 66, 120 V 6.6, 20, 66, 120 V 6.6, 20, 66, 120 V 6.6, 20, 66,120 V 6.6, 20, 66, 120 V 6.6, 20, 66, 120 V 6.6, 20, 66, 120 V
6.6, 20, 66, 120 V
Accuracy
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
0.05%+0.05%
Resolution
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Constant Power
Range
0 - 1 kW
0 - 2 kW
0 - 3 kW
0 - 6 kW
0 - 12 kW
0 - 18 kW
0 - 24 kW
0 - 36 kW
Accuracy
1% + 1%
1% + 1%
1% + 1%
1% + 1%
1% + 1%
1% + 1%
1% + 1%
1% + 1%
Resolution
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Constant Resistance
Range
5 m
⍀
- 180
⍀
2.5 m
⍀
- 90
⍀
1.67 m
⍀
- 60
⍀
833 µ
⍀
- 30
⍀
417 µ
⍀
- 15
⍀
278 µ
⍀
- 10
⍀
208 µ
⍀
-7.5
⍀
136 µ
⍀
- 5
⍀
Accuracy
5
2%
2%
2%
2%
2%
2%
2%
2%
Slew Rate (10 - 90%)
Range
1 A/s - 20 A/µs
2 A/s - 40 A/µs
3 A/s - 60 A/µs
6 A/s - 120 A/µs 12 A/s - 240 A/µs 18 A/s - 360 A/µs 24 A/s - 480 A/µs 36 A/s - 720 A/µs
Rise Time
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
10 µs - 20 s
Resolution
< 5 µs
< 5 µs
< 5 µs
< 5 µs
< 5 µs
< 5 µs
< 5 µs
< 5 µs
Accuracy
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
1% +/- 5µs
Short Circuit
Resistance
50, 5 m
⍀
25, 2.5 m
⍀
17, 1.7 m
⍀
8.3 m
⍀
- 833 µ
⍀
4.17 m
⍀
- 417 µ
⍀
2.78 m
⍀
- 278 µ
⍀
2.08 m
⍀
- 208 µ
⍀
1.39 m
⍀
- 139 µ
⍀
Current Max
33, 333 A
67, 667 A
60, 608 A
120, 1200 A
240, 2400 A
360, 3600 A
480, 4800 A
720, 7200 A
Macro
Modes
Any single Mode
Repetition
Single Burst or Continuous
Settings
100
Period
40 µs - 20 s
Delay
20 µs - 20 s
Resolution
10 µs
Accuracy
1% +/- 5 µs
Measurements
Accuracies: % of Measurement + % of Range, Resolution: % of Range
Current
Ranges
20, 200 A
40, 400 A
60, 600 A
120, 1200 A
240, 2400 A
360, 3600 A
480, 4800 A
720, 7200 A
Accuracy
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
0.12%+0.06%
Resolution
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
DC Voltage
Ranges
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
6.6, 66, 166 V
Accuracy
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
0.01%+0.02%
Resolution
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
0.0015%
Power
Ranges
Current Range x Voltage Range
Accuracy
Current Accuracy + Voltage Accuracy
Resolution
0.0015% Range
Waveform Capture
Bandwidth
25 kHz
Accuracy 1%
R
Channels
Voltage, Current or both MUX’d
Digitizing Rate
6
100 - 100K Samples/s
Memory
16K Samples
Timebase
10 µs - 8 s
Triggering
System or External
Waveform Analysis
Voltage, Current, Power, Overshoot, Undershoot, Rise/Fall Time, Turn-On Time, Settling Time, Hold-Up Time, AC RMS, AC+DC RMS
Control
User Interface
PC soft panel or manual touch-panel
PC Required
3 GHz µP with 512 MB RAM, SVGA display, 80 GB HD
OS
Windows XP, Vista
Test Executive
NI LabVIEW, emPower™ with integrated datalog/test report support
Communications
Ethernet (LXI), RS232, NHR RS485
Drivers
Active X
Physical
Load Connectors
Bus bars with lugs
Operating Temperature
0 - 40º C at full power and <75% duty cycle
Input Power
115/230 ± 10% VAC, 47 - 63 Hz
Dimensions (HxWxD)
5
1
/
4
x 19 x 22 in
5
1
/
4
x 19 x 22 in 10
1
/
2
x 19 x 22 in 10
1
/
2
x 19 x 22 in 35 x 23 x 30 in
43 x 23 x 30 in
57 x 23 x 30 in
72 x 23 x 30 in
Weight
40 lbs
50 lbs
75 lbs
100 lbs
250 lbs
400 lbs
570 lbs
815 lbs
Additional Features
Remote Voltage Sense
2 VDC max drop between sense and load input
Self Test
Power-up self test of all major functions including status of input, output, control and protection circuits
Performance Monitoring
Continuous checking of performance parameters and appropriate error messages and/or LED fault indicators when necessary
Calibration
Closed cover, all adjustments made in software and stored in EEPROM
Protection
OP, OC, OV, OT, Reverse Voltage and Undervoltage Lockout
Trigger Output
Synchronizes external device to programmed load step
Trigger Input
Synchronizes programmed load step to an external device
Current Monitor
0 - 10 V external signal appropriate to 100% current for the selected range
Analog Control
0 -10 V external signal appropriate to 100% current for the selected range
Fan Noise Reduction
Automatic fan speed control