Recorder modes – Atec Fluke-190 Series User Manual
Page 5
ScopeMeter 190 Series II Fluke Corporation 5
190-062
190-102
190-202
190-502
190-104
190-204
ScopeRecord
™
Roll Mode
Dual or multiple input waveform storage mode, using deep memory
Source and display
Input A, Input B, Dual.
All channels sampled simultaneously.
Any combination of inputs,
up to 4 channels. All channels
sampled simultaneously.
Bandwidth
20 MHz or 20 kHz, user selectable
Memory depth
30,000 data points, each holding min/max pair of information
Min/max values
Min/max values are created at samples that are measured at high sample rate
ensuring capture and display of glitches
Recording modes
Single sweep, continuous roll,
Start-on-Trigger (through external),
Stop-on-Trigger (through external)
Single sweep, continuous roll,
Start-on-Trigger (through any
channel), Stop-on-Trigger (through
any channel)
Stop-on-trigger
ScopeRecord mode can be stopped by an individual trigger event, or by an interruption
of a repetitive trigger signal, through any input channel (through External on 190-XX2 Series)
Horizontal scale
Time from start, time of day
Zoom
Ranges from full record overview to zoom in up to sample level, at any record length
Memory
Two multiple input ScopeRecord waveforms can be saved internally for later recall and analysis
Direct storage on external flash memory drive through USB host port
ScopeRecord
™
Roll mode sample rate and recording timespan
Time base range
5 ms/div ~ 2 min/div
Recorded timespan
6 sec ~ 48 hr
Time/division in ‘view all’ mode
0.5 s/div ~ 4 h/div
Glitch capture
8 ns
Sample rate
125 MS/s
Resolution
200 μsec ~ 4.8 sec
Trendplot
™
Recording
Multiple channel electronic paperless recorder. Graphically plots, displays and stores results of up to four automatic scope measurements or a
DMM-reading over time.
Source and display
Any combination of scope measurements, made on any of the input channels, or DMM reading
(2-channel instruments)
Memory depth
18,000 points (sets) per measurement. Each recorded sample point contains a minimum, a maximum and an
average value, plus a date- and timestamp.
Ranges
Normal view: 5 s/div to 30 min/div
In view-all mode: 5 min/div to 48 hr/div (overview of total record)
Recorded time span
Up to 22 days, with a resolution of 102 seconds
Recording mode
Continuous recording, starting at 5 s/div with automatic record compression
Measurement speed
3 automatic measurements per second or more
Horizontal scale
Time from start, time of day
Zoom
Up to 64x zoom-out for full record overview, up to 10x zoom-in for maximum detail
Memory
Two multiple input TrendPlot records can be saved internally for later recall and analysis
Direct storage on external flash memory drive through USB host port
Cursor measurements—all recorder modes
Source
Any waveform trace in any waveform display mode (Scope, ScopeRecord or TrendPlot)
Dual vertical lines
Cursors may be used to identify Min, Max or Average value of any datapoint in a record,
with time between cursors, time from start or absolute time.
Recorder modes