Electric and magnetic field measurement, Narda, Excellence – Atec Narda-EMR-300 User Manual
Page 6
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: [email protected] • www.narda-sts.com
GERMANY TEL: 49-7121-9732-777 • FAX: 49-7121-9732-790 • E-MAIL: [email protected] • www.narda-sts.de
Electric and Magnetic Field Measurement
50 Years of
Excellence
RF Safety
Test Solutions
an
communications company
narda
Safety Test Solutions
Linearity
c
Isotropic
Deviation
d
Overload
Limit
E-field
Suppression
Thermal
Response
e
Dimensions
Diameter,
Length
±dB (0.25 to 0.5 A/m)
±1 dB (0.5 to 1 A/m)
±0.5 dB (1 to 40 A/m)
±1 dB
(Typical, 100 kHz)
>500 A/m,
>5000 A/m
>20 dB, Typical
±0.8 dB
120 mm,
300 mm
±3 dB (0.017 to 0.033 A/m)
±1 dB (0.033 to 0.066 A/m)
±0.5 dB (0.066 to 3 A/m)
±1 dB (3 to 17 A/m)
±1 dB
(Typical, 27.12 MHz)
>35 A/m,
350 A/m
>20 dB, Typical
±0.8 dB
120 mm,
300 mm
±3 dB (0.025 to 0.05 A/m)
±1 dB (0.05 to 0.1 A/m)
±0.5 dB (0.1 to 3 A/m)
±1 dB (3 to 16 A/m)
±1 dB, Typical
20 A/m,
200 A/m
> 20 dB
±0.8 dB
65 mm,
310 mm
±1 dB (0.015 to 0.035 A/m)
±0.5 dB (0.035 to 1 A/m)
±1 dB (1 to 5 A/m)
±1 dB, Typical
> 6.3 A/m,
> 63 A/m
>20 dB, Typical
±.5/-0.8 dB
65 mm,
310 mm
Notes:
a
Typical performance data
b
Type 25 is shaped to FCC limits for "Controlled Occupational" environments,
Type 26 probe is shaped to Canadian Safety Code 6 (RF/microwave workers) and recommendations of ICNIRP for occupational exposures.
c
Linearity referred to 27.5 V/m at 27.12 MHz (27.5 V/m and 100 MHz for Type 9C, 27.5 V/m for Type 11C
and 50% of standard and 100 MHz for Types 25C and 26C) where there is an absolute error of ±1.0 dB
d
Includes EMR meter, Typical
e
0 to +50°C, except Types 11C, 25C and 26C which are -10 to +50°C