Electric and magnetic field measurement, Narda, Excellence – Atec Narda-EMR-300 User Manual
Page 4
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: [email protected] • www.narda-sts.com
GERMANY TEL: 49-7121-9732-777 • FAX: 49-7121-9732-790 • E-MAIL: [email protected] • www.narda-sts.de
Electric and Magnetic Field Measurement
50 Years of
Excellence
RF Safety
Test Solutions
an
communications company
narda
Safety Test Solutions
34
Linearity
c
Isotropic
Deviation
d
Overload
Limit
H-field
Suppression
Thermal
Response
e
Dimensions
Diameter,
Length
±3 dB (0.6 to 1.25 V/m)
±1.0 dB (1.25 to 2.5 V/m)
±0.5 dB (2.5 to 400 V/m)
±0.7 dB (400 to 800 V/m)
±1.0 dB for
f > 1 MHz
700 mW/cm
2
,
70 W/cm
2
>45 dB (300 kHz)
>35 dB (1 MHz)
>20 dB
(above 5 MHz)
+0.2 / -1.5 dB
64 mm,
310 mm
±0.5 dB (1.2 to 200 V/m)
±0.7 dB (200 to 320 V/m)
±1.0 dB
175 mW/cm
2
,
17.5 W/cm
2
>20 dB (Typ.)
+0.2 / -1.5 dB
75 mm,
310 mm
±3dB (0.8 to 1.65 V/m)
±1.0 dB (1.65 to 3.3 V/m)
±0.5 dB (3.3 to 300 V/m)
±8 dB (300 to 1000 V/m)
±1.5 dB
(10 MHz to 8 GHz)
±2.0 dB
(f > 8 GHz)
700 mW/cm
2
,
70 W/cm
2
> 20 dB
±0.8 dB
64 mm,
310 mm
+2/-3 dB (1.0 to 2.0 V/m)
±1.0 dB (2 to 250 V/m)
±1.0 dB
1600 V/m,
1 W/cm
2
>20 dB (Typ.)
±0.8 dB
64 mm,
310 mm
±3 dB (0.3 to 1.3%)
±1.0 dB (1.3 to 5%)
±0.5 dB (5 to 10,000%)
+3.0 dB/- 0.5 dB
(10 to 300 MHz)
±1.0 dB
(300 MHz to 1 GHz)
±2.0 dB
(1 to 12 GHz)
32 dB
(< 10kV/m),
50 dB
(< 100 kV/m)
>45 dB (300 kHz)
>35 dB (1 MHz)
>20 dB
(above 5 MHz)
+0.8 /- 1.0 dB
64 mm,
310 mm
±0.5 dB (39 to 614 V/m)
±1.0 dB
0.6W/m
2
,
200 W/m
2
N/A
±0 dB
64 mm,
310 mm
Notes:
a
Typical performance data
b
Type 25 is shaped to FCC limits for "Controlled Occupational" environments,
Type 26 probe is shaped to Canadian Safety Code 6 (RF/microwave workers) and recommendations of ICNIRP for occupational exposures.
c
Linearity referred to 27.5 V/m at 27.12 MHz (27.5 V/m and 100 MHz for Type 9C, 27.5 V/m for Type 11C
and 50% of standard and 100 MHz for Types 25C and 26C) where there is an absolute error of ±1.0 dB
d
Includes EMR meter, Typical
e
0 to +50°C, except Types 11C, 25C and 26C which are -10 to +50°C