Specifications

Radiation detected

Alpha

> 7.5 MeV

Beta

> 100 keV

Gamma

> 7 keV

Operating ranges,

response time

0 mR/h to 5 mR/h (8 sec)

0 mR/h to 50 mR/h (2.5 sec)

0 mR/h to 500 mR/h (2 sec)

0 R/h to 5 R/h (2 sec)

0 R/h to 50 R/h (2 sec)

Accuracy

Within 10 % of readings between 10 % and 100 % of full scale indication on any range,

exclusive of energy response

Detector

Chamber (cc volume air

ionization)

349 cc

Chamber wall (phenolic)

246 mg/cm

2

Chamber window (mylar)

6.6 mg/cm

2

Beta slide

440 mg/cm

2

Automatic features

Auto-zeroing, auto-ranging, and auto-backlight

Warm-up time

One minute

Display LCD analog/

digital with backlight

Analog

100 element bar graph 6.4 cm long. Bar graph is divided

into 5 major segments, each labeled with the appropriate

value for the range of the instrument

Digital

2.5 digit display is followed by a significant zero digit

depending on the operating range of the instrument. The

units of measurement are indicated on the display at all

times. Digits are 6.4 mm (0.25 in) high. Low battery and

freeze indicators are also provided on the display

Modes

Integrate mode

Operates continuously 30 seconds after the instrument

has been turned on. Integration is performed even if the

instrument is displaying in mR/h or R/h

Freeze mode

Will place a tick mark on the bar graph display to hold on

the peak displayed value. The unit will continue to read

and display current radiation values

Environmental

Temperature range

-20 °C to 70 °C (-4 °F to 158 °F)

Relative humidity

0 % to 100 % (at 140 °F)

Geotropism

< 1 %

Typical energy

dependence

16

Nitrogen gamma rays are 110 % to 120 % of indicated readings as determined at the

University of Lowell

Power requirements

Two 9 V alkaline, 200 hours operation

Dimensions (WxDxH)

10 cm x 20 cm x 15 cm (4 in x 8 in x 6 in)

Weight

1.11 kg (2.5 lb)

2 Fluke Biomedical 451B Ion Chamber Survey Meter with Beta Slide