Compatible with the Following Inspection Codes:
The OmniScan MX2 is compatible with standard industry inspection codes, including, but not limited to:
ASME Section V, Article 4
AWS
All ASME phased array code cases
API 1104 and API RP2X
ASTM E2700-09
CEN EN 583-6
ASTM E2491-06
BSI BS7706
… and more
OmniScan MX2 Mainframe Specifications
Overall dimensions
(W x H x D)
325 mm x 235 mm x 130 mm
(12.8 in. x 9.3 in. x 5.1 in.)
Weight
5 kg (11 lb), including module and one battery
Data Storage
Storage devices
SDHC card, most standard USB storage devices,
or fast Ethernet
Data file size
300 MB
I/O Ports
USB ports
3
Speaker out
Yes
Video output
Video out (SVGA)
Ethernet
10/100 Mbps
I/O Lines
Encoder
2-axis encoder line (quadrature, up, down, or
clock/direction)
Digital input
4 digital TTL inputs, 5 V
Digital output
4 digital TTL outputs, 5 V, 15 mA
Acquisition on/off
switch
Remote acquisition enabled TTL, 5 V
Power output line
5 V, 500 mA power output line
(short-circuit protected)
Alarms
3 TTL, 5 V, 15 mA
Analog output
2 analog outputs (12 bits) ±5 V in 10 kΩ
Pace input
5 V TTL pace input
Display
Display size
26.4 cm (10.4 in.) (diagonal)
Resolution
800 pixels x 600 pixels
Brightness
700 cd/m
2
Number of colors
16 million
Type
TFT LCD
Power Supply
Battery type
Smart Li-ion battery
Number of batteries
1 or 2 (battery chamber accommodates two
hot-swappable batteries)
Battery life
Minimum 6 hours with two batteries
Environmental Specifications
Operating temperature
range
0 °C to 45 °C; 0 °C to 35 ºC with 32:128 PA
(32 ºF to 113 ºF; 32 ºF to 95 ºF with 32:128 PA)
Storage temperature
range
–20 °C to 60 °C (–4 ºF to 140 ºF) with batteries
–20 °C to 70 °C (–4 ºF to 158 ºF) without bat-
teries
Relative humidity
0% to 85% noncondensing
No air intake; designed for IP66
Shockproof rating
Drop-tested according to MIL-STD-810G 516.6
Phased Array Module Specifications
(Applies to OMNI-M-PA32128)
Overall dimensions
(W x H x D)
244 mm x 182 mm x 57 mm
(9.6 in. x 7.1 in. x 2.1 in.)
Weight
1.2 kg (2.6 lb)
Connectors
1 OmniScan connector for phased array probes
Number of focal laws
256
Probe recognition
Automatic probe recognition
Pulser/Receiver
Aperture
32 elements
Number of elements
128 elements
Pulser
Voltage
45 V or 90 V per element
Pulse width
Adjustable from 30 ns to 500 ns, resolution of 2.5 ns
Pulse shape
Negative square wave
Output impedance
Less than 30 Ω
Receiver
Gain
0 dB to 74 dB, maximum input signal 1.25 Vp-p
Input impedance
50 Ω
System bandwidth
0.53 MHz to 21 MHz (–3 dB)
Beamforming
Scan type
Sectorial and linear
Group quantity
Up to 8
Active elements
32
Elements
128
Data Acquisition
Digitizing frequency
100 MHz (10 bits)
Maximum pulsing rate
Up to 10 kHz (C-scan)
Data Processing
Number of data points
Up to 8,000
Real-time averaging
2, 4, 8, 16
Rectifier
RF, full wave, halfwave +, halfwave –
Filtering
Low-pass (adjusted to probe frequency), digital filtering
(bandwidth, frequency range)
Video filtering
Smoothing (adjusted to probe frequency range)
Data Visualization
A-scan refresh rate
Real time: 60 Hz
Data Synchronization
On internal clock
1 Hz to 10 kHz
On encoder
On 1 or 2 axes
Programmable Time-Corrected Gain (TCG)
Number of points
16 (1 TCG curve per channel for focal laws)
Alarms
Number of alarms
3
Conditions
Any logical combination of gates
Analog outputs
2
OmniScan MX2 Specifications