beautypg.com

02_en, Noise laboratory co., ltd, Option – Atec Noiseken-LSS-6230 User Manual

Page 2: Cdn for interconnection lines model:lss-inj6400sig, Isolation transformer model:tf-2302p

background image

Noise Laboratory Co., Ltd.

1-4-4 Chiyoda, Chuo-ku, Sagamihara City, Kanagawa Pref. 252-0237 Japan

International Sales & Marketing Section

TEL : +81-(0)42-712-2051/FAX : +81-(0)42-712-2050

E-mail : [email protected]

URL : http://www.noiseken.com

Terminal Connection Board attached with Multi-Outlet MODEL:18-00048B/58B

Enable to connect any outlet figure in the world.

18-00048B : For single phase 3 lines

(Voltage strength 4.5kV)

18-00058B : For 3-phase 5 lines

(Voltage strength 4.5kV)

Optical USB module MODEL:07-00022A

Optical conversion adaptor Used for remote
control with PC. 5m of optical fiber cable with
USB interface attached.

Warning lamp MODEL:11-00008A

CDN for Telecom lines MODEL:LSS-INJ6401TEL

The blinking makes the operators or neighbors pay
attention to the test processing.

EUT Protective Safety Box MODEL:11-00005A/11-00006A

Protection box to prevent access to EUT during the test.
Further safety is secured together with the safety
protective fence

CDN for interconnection lines MODEL:LSS-INJ6400SIG

Adaptable for the test to interconnection lines
specified in IEC61000-4-5 Standard. Arrestor
installation possible in between the each line and
GND.

* Please contact us if low voltage surge injection is required

Parameter

Specification

Surge input voltage

500V∼6,600V

EUT power capacity

DC50V/1A

Max. line number

4 lines

Decoupling coil

20mH per phase

Matching resistor

40Ω±10%

Dimension / Mass

W488×H456×D550mm Approx. 45kg

CDN for 3-phase EUT MODEL:LSS-CDN6351

Adaptable for the surge immunity test specified in IEC61000-4-5 Standard to AC power supply lines of EUT whose power supply up to
AC500V/50A. The surge injection angle can be set automatically linked with the generator unit. Also, the line symmetrical operation to AC
lines is possible.

* Scheduling to be released

* Photo in the development stage

* Photo in the development stage

Parameter

Specification

Coupling surge waveform

1.2/50μs-8/20μs combination waveforms, 0.5μs-100kHz ring wave

Max. coupling surge voltage / current

Up to the value which can be set in LSS-6230

Coupling network

18μF±10%(LINE - LINE)

Correspondent to IEC 61000-4-5

10Ω+9μF±10%(LINE - PE)

Coupling network

18μF±10%(LINE - LINE, LINE - PE)

ICorrespondent to IEEE/ANSI C62.45

9μF/9μF/9μF/9μF±10%(4 Lines - PE simultaneous injection)

Power supply lines structure for EUT

3-phase : L1/L2/L3/N/PE

EUT power capacity

AC500V/50A MAX 50/60Hz

Decoupling coil

1.5mH

Power supply

AC100V∼AC240V ±10% 50Hz / 60Hz

Dimension / Mass

W555×H950×D790 mm(Projection excluded) Approx. 120kg(To be fixed)

Isolation transformer MODEL:TF-2302P

Versatilely usable not only for AC power supply of our
lightning surge simulators but also for kinds of noise
testes and measurement equipments.

Adaptable for the test to unshielded symmetrical
interconnection lines specified in IEC61000-4-5
Standard

Parameter

Specification

Max. input voltage

Single phase AC 240V MAX (50/60Hz)

Max. output voltage 30A MAX

Dielectric strength

Primary to core : AC4kV (1 minute)

Secondary to core : AC4kV (1 minute)

Primary-Secondary : AC4kV (1 minute)

Insulation resistance ≧100MΩ at DC500V

Dimension / Mass

W350×H475×D 400mm

(Eye bolts and handles excluded) Approx. 60 kg

Parameter

Specification

Surge input voltage

6.6kV 10/700μs-5/320μs Combination waveform

EUT power capacity DC50V 100mA

No. of line

4 lines

Decoupling coil

20mH per phase

Matching resistor

40Ω

(in combination waveforms)

25Ω

(in 10/700μs waveform)

Dimension / Mass

W297×H262×D250mm Approx. 10kg

Loop coil MODEL:01-00057A

In the combination with LSS-6230, enable to
conduct pulse magnetic field immunity test
specified in IEC61000-4-9 Standard

* Scheduling to be released

Parameter

Specification

Max. output voltage 1200A

Waveform

6.4/16μAcurrent surge

Coil dimension

1000×1000mm

MODEL

Dimension

11-00005A

W400×D300×H300mm

11-00006A

W600×D350×H400mm

Option

Designs and specifications are subject to be changed without notice.