02_en, Noise laboratory co., ltd, Option – Atec Noiseken-LSS-6230 User Manual
Page 2: Cdn for interconnection lines model:lss-inj6400sig, Isolation transformer model:tf-2302p
Noise Laboratory Co., Ltd.
1-4-4 Chiyoda, Chuo-ku, Sagamihara City, Kanagawa Pref. 252-0237 Japan
●
International Sales & Marketing Section
TEL : +81-(0)42-712-2051/FAX : +81-(0)42-712-2050
E-mail : [email protected]
URL : http://www.noiseken.com
Terminal Connection Board attached with Multi-Outlet MODEL:18-00048B/58B
Enable to connect any outlet figure in the world.
18-00048B : For single phase 3 lines
(Voltage strength 4.5kV)
18-00058B : For 3-phase 5 lines
(Voltage strength 4.5kV)
Optical USB module MODEL:07-00022A
Optical conversion adaptor Used for remote
control with PC. 5m of optical fiber cable with
USB interface attached.
Warning lamp MODEL:11-00008A
CDN for Telecom lines MODEL:LSS-INJ6401TEL
The blinking makes the operators or neighbors pay
attention to the test processing.
EUT Protective Safety Box MODEL:11-00005A/11-00006A
Protection box to prevent access to EUT during the test.
Further safety is secured together with the safety
protective fence
CDN for interconnection lines MODEL:LSS-INJ6400SIG
Adaptable for the test to interconnection lines
specified in IEC61000-4-5 Standard. Arrestor
installation possible in between the each line and
GND.
* Please contact us if low voltage surge injection is required
Parameter
Specification
Surge input voltage
500V∼6,600V
EUT power capacity
DC50V/1A
Max. line number
4 lines
Decoupling coil
20mH per phase
Matching resistor
40Ω±10%
Dimension / Mass
W488×H456×D550mm Approx. 45kg
CDN for 3-phase EUT MODEL:LSS-CDN6351
Adaptable for the surge immunity test specified in IEC61000-4-5 Standard to AC power supply lines of EUT whose power supply up to
AC500V/50A. The surge injection angle can be set automatically linked with the generator unit. Also, the line symmetrical operation to AC
lines is possible.
* Scheduling to be released
* Photo in the development stage
* Photo in the development stage
Parameter
Specification
Coupling surge waveform
1.2/50μs-8/20μs combination waveforms, 0.5μs-100kHz ring wave
Max. coupling surge voltage / current
Up to the value which can be set in LSS-6230
Coupling network
18μF±10%(LINE - LINE)
Correspondent to IEC 61000-4-5
10Ω+9μF±10%(LINE - PE)
Coupling network
18μF±10%(LINE - LINE, LINE - PE)
ICorrespondent to IEEE/ANSI C62.45
9μF/9μF/9μF/9μF±10%(4 Lines - PE simultaneous injection)
Power supply lines structure for EUT
3-phase : L1/L2/L3/N/PE
EUT power capacity
AC500V/50A MAX 50/60Hz
Decoupling coil
1.5mH
Power supply
AC100V∼AC240V ±10% 50Hz / 60Hz
Dimension / Mass
W555×H950×D790 mm(Projection excluded) Approx. 120kg(To be fixed)
Isolation transformer MODEL:TF-2302P
Versatilely usable not only for AC power supply of our
lightning surge simulators but also for kinds of noise
testes and measurement equipments.
Adaptable for the test to unshielded symmetrical
interconnection lines specified in IEC61000-4-5
Standard
Parameter
Specification
Max. input voltage
Single phase AC 240V MAX (50/60Hz)
Max. output voltage 30A MAX
Dielectric strength
Primary to core : AC4kV (1 minute)
Secondary to core : AC4kV (1 minute)
Primary-Secondary : AC4kV (1 minute)
Insulation resistance ≧100MΩ at DC500V
Dimension / Mass
W350×H475×D 400mm
(Eye bolts and handles excluded) Approx. 60 kg
Parameter
Specification
Surge input voltage
6.6kV 10/700μs-5/320μs Combination waveform
EUT power capacity DC50V 100mA
No. of line
4 lines
Decoupling coil
20mH per phase
Matching resistor
40Ω
(in combination waveforms)
25Ω
(in 10/700μs waveform)
Dimension / Mass
W297×H262×D250mm Approx. 10kg
Loop coil MODEL:01-00057A
In the combination with LSS-6230, enable to
conduct pulse magnetic field immunity test
specified in IEC61000-4-9 Standard
* Scheduling to be released
Parameter
Specification
Max. output voltage 1200A
Waveform
6.4/16μAcurrent surge
Coil dimension
1000×1000mm
MODEL
Dimension
11-00005A
W400×D300×H300mm
11-00006A
W600×D350×H400mm
Option