T e c h n ic a l S p ec if i c a t io ns
Wave Shape into 50Ω load
Rise time
5 ns ± 30 %
Impulse duration
50 ns ± 30 %
Wave Shape into 1000Ω load
Rise time
5 ns ± 30 %
Impulse duration
50 ns – 15 / + 100 ns
Burst Characteristic
Voltage at HV-output:
Voltage at CDN output:
1.0 … 7.3 kV
1.0 … 6.6 kV
Spike frequency
1 Hz … 110 kHz
Burst duration
0.01 … 999 ms
Burst period
0.1 Hz … 400 Hz
2.5 ms … 10 s
1 period … 500 periods
Impulses per second
max. 250 for Unom > 4 kV
max. 500 for Unom
≤
4 kV
Impulses per burst
max. 75 for Unom > 4 kV
max. 150 for Unom
≤
4 kV
Spike distribution
normal, continuous,
random, real
Polarity
positive, negative
Test time
10 s … 8 h per path
Weight and Dimension
Weight
14 kg
W x H x D
450mm x 130mm x 570mm
Single Phase Coupling/Decoupling Network
Maximum AC voltage
264 V @ 16 … 440 Hz
Maximum AC current
16 A @ 50 … 60 Hz, 10 A @
400 Hz
Maximum DC voltage
125 V
Maximum DC current
16 A (dependant on DC
voltage)
Coupling Modes
L-GND, N-GND, PE-GND,
LN-GND, LPE-GND, NPE-
GND, LNPE-GND
Residual voltage at
Test Supply Input
≤ 10% of applied
test voltage
Synchronisation
Frequency
16
2
/
3
Hz, 40 Hz, 50 Hz,
60 Hz, 400 Hz, auto
Impulse trigger
automatic, manual, external
System Configuration
EUT fail input
BNC, logic low to trigger
P90 interface
for a 3-phase CDN
extension
PESD interface
for connecting an ESD
generator
Printer interface
Centronix
Computer interface
RS-232, IEEE-488 (optional)
Power Supply
Voltage
85 … 264 VAC, 150 VA
Frequency
50 / 60 Hz
Impulse duration
burst duration
burst period
t
t
V
V
90%
50%
10%
rise time
spike frequency