Control Computer

The controller is a standard PC with operating system Microsoft

®

Windows. Depending on the system layout GPIB (IEEE488) with
National Instruments interface card, serial bus RS232, USB and other
bus systems are supported.

Control Software

CR-LAB is fully compliant to IEC/EN 61000-4-3, IEC/EN 61000-4-20,
automotive and MIL standards. The software CR-LAB controls
the complete test system and creates the test report. It performs
measurement of the uniform area and generates reference calibration
data from it. Alternatively reference data can be measured directly.
Tests may be performed manually and fully automatic. A full
automatic monitoring of the EUT´s (Equipment Under Test) function
is possible whenever its compliance can be controlled with preset
tolerance limits. Up to four values can be monitored and recorded for
example by means of multimeters.

Signal Generator

As signal source a commercial signal generator or the built-in
signal generator of our EMC test & control unit ECU-3/-6 is
used. It should cover at least the required frequency range
such as 80 MHz to 6 GHz and allow amplitude modulation
with a sine wave of 1 kHz and 80 %, as requested by the
standards. Besides, it should meet the requirements regarding
frequency step width (1 % of the preceding value). If the
equipment offers further modulation depth and modes,
as well as a higher frequency range and smaller steps, this
might be advantageous for future applications. Minimum
requirements are as following:

Essential data of the software are:

• Microsoft

®

Windows platform

• Simple operator´s guide
• Online help function
• Presentation of the results in online graphics and reports
• Export function of the files for further processing under
Microsoft

®

Word, Microsoft

®

Excel,…

• Measurement of homogeneous field incl. evaluation
• Calculation of reference data
• 2dB saturation test on base of homogeneous field
measurement compliant to standard
• Measurement of reference data with fixed test level or
profile of level vs. frequency
• Permanent VSWR control during test and operator
defined limitation as well as restriction of max.
input level of amplifier and max. allowed output power
• Automatic multiple repetition of test
• Manual test mode
• Manual increase/decrease of test level
• Automatic test mode incl. monitoring of the EUT
• Handshake function to EUT via serial interface
• Easy and fast graphical device set-up, system layout

can be printed

• Fully compliant to IEC/EN 61000-4-3, IEC/EN 61000-4-20,

automotive- and MIL standards

• Control of the test system by GPIB, USB interfaces,…
• Customized modifications possible

Frequency range: 9 kHz – 3.0 GHz (6.0 GHz)

RF output: -40 dBm to 0 dBm

Frequency resolution: 1 Hz

Level resolution: 0.1 dB

Amplitude modulation: 0 to 99.9 %

Further modulation types: frequency modulation,

phase modulation, pulse modulation

Interface: GPIB (IEEE-488), RS232, USB

Immunity Test System RIS 3000 / RIS 6000

acc. to IEC/EN 61000-4-3

5 // PAGE