Atec Haefely-Axos5 User Manual
Page 5
5
Compact Immunity Test System
ICE / EN 61000-4-5 EDITION 2 SURGE COMBINATION WAVE
Output voltage
0.2 – 5.0 kV ±10%
Output current
0.1 – 2.5 kA ±10%
Voltage rise time
1.2 us ±30%
Current rise time
8 us ±20%
Voltage duration
50 us ±20%
Current duration
20 us ±20%
Polarity
pos / neg / alternate
Integrated single
phase CDN
264 V AC / 16 A
220 V DC / 10 A
Output impedance
2 Ohms
Phase synch
0 – 359° with 1° steps or
asynchronous mode
Impulse trigger
automatic 5 s – 99999 s
manual
external trigger input
Counter preselect
counter
1 – 10000 / infinite
100000
Peak voltage
monitor
BNC output: 1000:1
display: 3 digits
Peak current
monitor
BNC output: 1 kA/V
display: 3 digits
IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS
Max. voltage
264 V AC/DC
Interrupt time
0.1 period – 9999 periods
0.02 ms – 999 minutes
Max. current
16 A AC/DC continuous
20 A for 5 s
40 A for 3 s
500 A inrush current
Interval time
1 period – 9999 periods
synch
0.04 ms – 999 minutes
asynch
Trigger
automatic
manual
external trigger input
Test time
1 s – 999 minutes
infinite
Interrupt
dip level
0%
0% – 99% with external
voltage source
Phase synch
0 – 359° 16 / 40 / 50 / 60
Hz
asynchronous mode
RMS voltage
monitor
BNC output: 100:1
display: 4 digits
RMS current
monitor
BNC output: 10 A/V
display: 4 digits