Compact Immunity Test System

ICE / EN 61000-4-5 EDITION 2 SURGE COMBINATION WAVE
Output voltage

0.2 – 5.0 kV ±10%

Output current

0.1 – 2.5 kA ±10%

Voltage rise time

1.2 us ±30%

Current rise time

8 us ±20%

Voltage duration

50 us ±20%

Current duration

20 us ±20%

Polarity

pos / neg / alternate

Integrated single
phase CDN

264 V AC / 16 A
220 V DC / 10 A

Output impedance

2 Ohms

Phase synch

0 – 359° with 1° steps or
asynchronous mode

Impulse trigger

automatic 5 s – 99999 s
manual
external trigger input

Counter preselect
counter

1 – 10000 / infinite
100000

Peak voltage
monitor

BNC output: 1000:1
display: 3 digits

Peak current
monitor

BNC output: 1 kA/V
display: 3 digits

IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS
Max. voltage

264 V AC/DC

Interrupt time

0.1 period – 9999 periods
0.02 ms – 999 minutes

Max. current

16 A AC/DC continuous
20 A for 5 s
40 A for 3 s
500 A inrush current

Interval time

1 period – 9999 periods
synch
0.04 ms – 999 minutes
asynch

Trigger

automatic
manual
external trigger input

Test time

1 s – 999 minutes
infinite

Interrupt
dip level

0%
0% – 99% with external
voltage source

Phase synch

0 – 359° 16 / 40 / 50 / 60
Hz
asynchronous mode

RMS voltage
monitor

BNC output: 100:1
display: 4 digits

RMS current
monitor

BNC output: 10 A/V
display: 4 digits