L2-16b, Related products, Ordering information – Atec PMM-L2-16B User Manual
Page 2: Optional accessories, L2-16b equivalent circuit
L2-16B equivalent circuit
L2-16B
Two Lines Single Phase V-Network 9 kHz - 30 MHz, 16 A for AC and DC powered EUT
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23
20090 Segrate (Milano) - ITALY
Phone: +39 02 2699871
Fax: +39 02 26998700
E-Mail: [email protected]
Internet: www.narda-sts.it
L216B-FEN-30400 - Specifications subject to changes without prior notice
•
Electrical safety and presence
of ground protection relays do
require the installation of properly
rated insulating transformer(s)
between mains power line and
AMN line inputs.
•
High mains noise may require
the installation of properly rated
mains filters to reduce the level of
unwanted signals.
9 kHz - 30 MHz
16 A
250 VAC
350 VDC
DC - 60 Hz
50 Ω // [5 Ω + 50 µH]
with 250µH choke
BNC female
SCHUKO connector
-10 °C to +45 °C
-25 °C to +75 °C
230 x 105 x 285 mm
5,5 kg
SPECIFICATIONS
Frequency range
Continuous rated output current
Maximum permissible operating
voltage (L/N) L/PE)
AC supply frequency range
Equivalent circuit
RF output
Test item
Rated temperature
Storage temperature
Overall Dimensions mm (W x H x D)
Weight
Receivers
LISN
RFI Filters
• 7010: EMI receiver 150 kHz -1 GHz
• 9010: EMI receiver 10 Hz - 30 MHz
• 9010F: EMI receiver 10 Hz - 30 MHz
• 9010/03P: EMI receiver 10 Hz - 300 MHz
• 9010/30P: EMI receiver 10 Hz - 3 GHz
• 9010/60P: EMI receiver 10 Hz - 6 GHz
• L3-32: 4 lines, 3-phase AMN, 32 A
• L3-64: 4 lines, 3-phase AMN, 64 A
• L3-64/690 4 lines, 3-phase AMN, 64A/690V
• L3-100: 4 lines, 3-phase AMN, 100 A
• L3-500: 4 lines, 3-phase AMN, 350 A
• L1-150M: single-path, 50 Ohm etc AMN 150 A
• L1-500 Single phase AMN, 500A
• L2-D: Delta LISN for telecom, 2 A, 150 Ω
• FIL-L2-16F: single phase RFI filter, 16 A
• FIL-L2-24M: single phase RFI filter, 24 A
• FIL-L3-32M: 3-phase+neutral RFI filter, 32 A
• FIL-L3-70M: 3-phase+neutral RFI filter, 70 A
Ordering Information:
L2-16B single phase Artificial Mains Network
Includes:
power supply cable, RF cable, LISN remote control cable,
user’s manual, calibration certificate.
Optional accessories:
LISN service kit
(AC-BNC adapter for LISN verification and calibration)
Related Products