Atec Megger-BM Series User Manual
Page 2
BM11
BM11D
Features
n
Large direct reading analogue scale
n
Large Analogue/digital display
n
Guard terminal
n
Built-in timer
n
Rugged case design
n
Locking test leads for safety
n
123 mm scale length
n
Battery indicator
n
Weatherproof to IP54
Note: Not for sale in E.U.
n
Voltage measurement and warning
BM11, BM11D, BM21 and BM25
5 kV Analogue/Digital Insulation Testers
FeATUres AnD BeneFITs
All BM11 series
n
Choice of test voltages
n
Direct reading display
n
Unique rugged case
n
Guard terminal
n
Designed for user safety
BM11D
n
Large, custom Analogue/Digital LCD
n
Built-in timer
n
Locking test leads for safety
n
Battery indicator
n
Automatic discharge of capacitance
n
Voltage measurement and warning
n
Tests to 500 GΩ
n
Weather proof to IP54
Test voltages (d.c.) 500, 1000, 2500, 5000 V
500, 1000, 2500, 5000 V
Accuracy (20°C)
±5% on open circuit
±5% on 100 MΩ load
Insulation
100 kΩ to 100 GΩ; at all test voltages
Digital 10 kΩ to 50 GΩ @ 500 V
Resistance
10 kΩ to 100 GΩ @ 1000 V
Range
10 kΩ to 250 GΩ @ 2500 V
10 kΩ to 500 GΩ @ 5 kV
Analogue 100 kΩ to 1 TΩ @ all voltages
Basic Accuracy
±1,5% of arc length
±5% of reading 10 MΩ to 100 GΩ @ 5 kV
Short Circuit
Current
1 mA nominal, 2 mA max.
1,8 mA nominal, 2 mA max.
Voltage Range
50 to 1000 V d.c. or a.c.
50 to 1000 V d.c. or a.c.
Accuracy (20°C)
±5% ±10 V
±2% ±1 V
Display
Analogue
Analogue/digital (3 digits)
Interference
Rejection
2 mA a.c. (50/60 Hz) on 5 kV range
1 mA r.m.s. per kV to a maximum 2 mA
Capacitor
Discharge Time
<1 s per µF to discharge from 5000 V to 50 V
<2 s per µF to discharge from 5000 V to 50 V
Leakage current
measurement
None
None
Capacitance
measurement
None
None
Timer
None
Automatic; 0 to 60 minutes
Temperature
Operating -15 to +55°C (5 to +131°F)
Operating -20to +50°C (-4 to 122°F)
Range
Storage -40 to +65 °C (-40 to +149°F)
Storage -25 to +65 °C (-13 to +149°F)
Temperature
Coefficient
< 0,04% of arc length per °C (0.2%/°F)
(Applies over range 0 to 30°C) (32 to 86°F)
0,2% per °C (0.1%/°F) for test currents >100 nA 0,1% per °C to test voltage (0.05% per °F)
Humidity Range
90% RH @ 40°C max. (104°F)
90% RH @ 40°C max. (104°F)
Safety
IEC348
IEC1010-1 (1995), EN61010 (1995) to installation
category III, 300 V, phase to earth, 500 V phase to phase