2  Fluke Corporation     Fluke 1630 Earth Ground Clamp Meter

Stakeless testing system

The Fluke 1630 uses the Stakeless testing method, 
which eliminates the need to disconnect parallel 
ground rods and find suitable locations for 
placing auxiliary ground stakes. This saves time 
and enables users such as industrial and utility 
electricians, and field service electricians and 
contractors, to perform earth ground loop tests 
in locations where it is not possible to use other 
techniques, such as inside buildings or on power 
pylons.  With the Stakeless testing method, earth 
ground stakes are no longer necessary.  The Fluke 
1630 Earth Ground Clamp is placed around the 
earth ground rod or the connecting cable. A known 
voltage is induced by one half of the clamp and 
the current is measured by the other half. The 
tester automatically determines the ground loop 

resistance at this grounding connection. 

Applications

  Earth Ground loop testing in various installations 

- for example, high voltage pylons, buildings, cell 

phone substations, RF transmitters, etc.

  Inspection of lightning protection systems

  Leakage current measurement

General specifications

Operating error

Refers to the operating temperature range and is guaranteed 

for 2 years

Storage temp. range

-20 °C to 60 °C (-4 °F to 140 °F ) 

Reference temp. range

23 °C ± 5 °C (73 °F ± 9 °F )

Storage humidity

< 75 % R

H

Operating humidity

< 85 % R

H

Display

9999 digit LC-display with special symbols

Protection

IP20 according to IEC 529/EN 60529

Safety

300 V, CAT III / pollution degree 2 IEC 61010-1

Weight

640 g (1.4 lb) 

Conductor size

35 mm (1.38 in) approximately 

Dimension (length x width x depth)

257 mm x 100 mm x 47 mm 

(1 in x 3.9 in x 1.9 in)

Emission

IEC 1000 4-2, IEC 61326-1 class B 

Immunity

IEC 61000-4-2, 8 kV (air) criteria A
IEC 61000-4-3, 3 V/m performance criteria A

Range selection

Auto

Overload indicator

“OL” on display

Measuring time

0.4 sec

Measuring frequency

3.333 kHz

Battery type

9 V alkaline (type IEC 6 LR 61)

Battery life

≥ 8 hours (continuous operation)

Power consumption

40 mA

Low battery indication

Display symbol