Weight

15.9 kg (35 lbs)

Dimensions (W x H x D, without handle) 450 x 145 x 390 mm (17.7 x 5.7 x 15.4 in)
PC connection

Two 10/100 Mbit/s PoE

4

Ethernet ports:

• 10/100 Mbit/s (10/100 Base-TX,

auto-sensing, auto-crossover, RJ45
connector for twisted pair cables)

• IEEE 802.3af compliant
• Port capability limited to one Class 1

(3.84 W) and one Class 2 (6.49 W)
powered device

If equipped with PAR-1 option: Parallel
port (IEEE 1284-C connector). Replaces
the standard Ethernet interfaces by a
parallel port if required by the used PC or
if the binary extension device CMB IO-7
should be used

5

Signal indication (LED)

> 42 V for voltage outputs and AUX DC

Connection to ground (earth)

4 mm (0.16 in) banana socket (rear side)

Hardware diagnostics

Self diagnostics upon each start-up

Galvanically separated groups

The following groups are galvanically
separated from each other: mains, volt-
age amplifier output, current amplifier
group A/B, auxiliary DC supply, binary/
analog input

Protection

All current and voltage outputs are fully
overload and short circuit proof and
protected against external high-voltage
transient signals and over temperature

Certifications

Developed and manufactured under an
ISO 9001 registered system

Technical Data CMC 256plus (continued)

IEC 61850 GOOSE

1

Simulation

Mapping of binary outputs to data attributes in
published GOOSE messages.
Number of virtual binary outputs: 360
Number of GOOSEs to be published: 128

Subscription

Mapping of data attributes from subscribed GOOSE
messages to binary inputs.
Number of virtual binary inputs: 360
Number of GOOSEs to be subscribed: 128

Performance

Type 1A; Class P2/3 (IEC 61850-5).
Processing time (application to network or vice versa):
< 1 ms

VLAN support

Selectable priority and VLAN-ID

IEC 61850 Sampled Values (Publishing)

1

Specification

According to the “Implementation Guideline for
Digital Interface to Instrument Transformers Using
IEC 61850-9-2“ of the UCA International Users Group

Sampling Rate

80 samples per cycle for nominal frequencies of 50
Hz and 60 Hz; synchronized with CMIRIG-B.

Synchronization

Synchronization attribute (smpSynch) is set when
the CMC is in synchronized operation mode utilizing
CMIRIG-B.
Sample count (smpCnt) zero is aligned with top of
the second (IRIG-B and PPS)
Accuracy data see above

VLAN support

Selectable priority and VLAN-ID

Max. number of SV streams 2 (with option LLO-2: 3 SV streams)
Power supply
Nominal input voltage

2

100 – 240 VAC, 1-phase

Permissible input voltage

85 ... 264 VAC

Nominal frequency

50/60 Hz

Permissible frequency range 45 ... 65 Hz
Power consumption

1.4 kVA at 115 V

/ 2.3 kVA at 230 V

Rated current

12 A at 115 V / 10 A at 230 V

Connection

Standard AC socket (IEC 60320)

Environmental conditions
Operation temperature

3

0 ... +50 °C (+32 ... +122 °F)

Storage temperature

-25 ... +70 °C (-13 ... +158 °F)

Humidity range

Relative humidity 5 ... 95 %, non-condensing

Vibration

IEC 60068-2-6 (20 m/s

2

at 10 ... 150 Hz)

Shock

IEC 60068-2-27 (15 g/11 ms half-sine)

Safety Standards, Electromagnetic Compatibility
EMC

The product adheres to the electromagnetic compat-
ibility (EMC) Directive 2004/108/EC (CE conform).

Emission

Europe EN 61326-1; EN 61000-6-4; EN 61000-3-2/3

International IEC 61326-1;IEC 61000-6-4; IEC 61000-3-2/3

USA FCC Subpart B of Part 15 Class A

Immunity

Europe EN 61326-1; EN 61000-6-2;

EN 61000-4-2/3/4/5/6/11

International IEC 61326-1;IEC 61000-6-2;

IEC 61000-4-2/3/4/5/6/11

Safety

The product adheres to the low voltage Directive
2006/95/EC (CE conform).

Europe EN 61010-1 Insulation of PC and SELV interfaces

complies with EN 60950-1

International IEC 61010-1

USA UL 61010-1

Canada CAN/CSA-C22.2 No 61010-1-04

1

Testing with GOOSE and Sampled Values functionality requires software licences for

the corresponding configuration modules

2

For line inputs below 115 VAC, it is not possible to drive all outputs (voltage

output, current output, Aux DC) simultaneously at full load. All other technical

specifications (e.g. the maximum output power of a single amplifier) are not

affected

3

For an operational temperature above +30 °C (+86 °F) a duty cycle of down to

50 % may apply

4

PoE = Power over Ethernet

5

CMC 256plus with PAR-1 option is connected to the USB port of the PC by means

of the parallelport cable and the CMUSB-P (USB to Parallel Port Converter).

CMC 256plus equipped with the PAR-1 option can not be extended with the LLO-2

option and does not support the CMIRIG-B accessory

Ordering Information

CMC 256plus with Test Universe Software

VE002701

CMC 256plus Basic

VE002702

CMC 256plus Protection

VE002703

CMC 256plus Advanced Protection

VE002704

CMC 256plus Universal

VE002705

CMC 256plus Meter

VE002706

CMC 256plus Measurement

CMC 256plus with CMControl (without Test Universe Software)

VE002715

CMC 256plus with CMControl-6

The CMControl can also be ordered as add-on together with a
CMC 256plus with Test Universe software or as a later upgrade.

CMC 256plus Hardware Options

VEHO2703

Option LLO-2 if ordered with a new unit

VEHO2704

Option LLO-2 if ordered as an upgrade

VEHO2701

Option PAR-1 if ordered with a new unit

5

VEHO2702

Option PAR-1 if ordered as an upgrade

5