beautypg.com

Atec OMICRON-CMC-256-Plus User Manual

Page 2

background image

Low level outputs

1

Number of outputs

6 (12 with Option LLO-2)

Setting range

0 ... ±10 Vpk

Max. output current

1 mA

Accuracy

error < 0.025 % typ., < 0.07 % guar.
at 1 ... 10 Vpk

Resolution

250 µV

Distortion (THD+N)

2

< 0.015 % typ., < 0.05 % guar.

Unconventional CT/VT simulation

linear, Rogowski (transient and sinewave)

Overload indication

yes

Isolation

SELV

Usability

completely independent from internal
amplifier outputs

Connection

16 pin combination socket (rear side)

Auxiliary DC supply
Voltage ranges

0 ... 264 VDC, 0.2 A / 0 ... 132 VDC,
0.4 A / 0 ... 66 VDC, 0.8 A

Power

max. 50 W

Accuracy

error < 2 % typ., < 5 % guar.

Connection

4 mm (0.16 in) banana sockets

Binary inputs
Number

10

Trigger criteria

Toggling of potential-free contacts or DC
voltage compared to threshold voltage

Input characteristics

0 ... ±600 VDC threshold or
potential-free

Ranges

100 mV / 1 V / 10 V / 100 V / 600 V

Resolution of threshold

±2 mV, ±20 mV, ±200 mV, ±2 V,
±20 V in ranges

Sample rate

10 kHz (resolution 100 µs)

Time stamping accuracy

±0.00015 % of rd.

3

±70 µs

Max. measuring time

infinite

Debounce / Deglitch time

0 ... 25 ms / 0 ... 25 ms

Counting function

< 3 kHz at pulse width > 150 µs

Galvanic isolation

5 galvanically isolated groups
(2+2+2+2+2)

Max. input voltage

CAT IV / 150 V, CAT III / 300V, CAT II /
600 V (850 Vpk)

Connection

4 mm (0.16 in) banana sockets

Counter inputs 100 kHz
Number

2

Max. counting frequency

100 kHz

Pulse width

> 3 µs

Threshold voltage

6 V

Voltage hysteresis

2 V

Max. input voltage

±30 V

Isolation

SELV

Connection

16 pin combination socket (rear side)

Trigger on overload
Supported generators

Current generators

Timer accuracy

error < 1 ms

Binary outputs, relays
Type

potential-free relay contacts, software
controlled

Number

4

Break capacity AC

Vmax: 300 VAC / Imax: 8 A /
Pmax: 2000 VA

Break capacity DC

Vmax: 300 VDC / Imax: 8 A /
Pmax: 50 W

Connection

4 mm (0.16 in) banana sockets

Binary outputs, transistor
Type

open collector transistor outputs

Number

4

Update rate

10 kHz

Imax

5 mA

Connection

16 pin combination socket (rear side)

DC voltage measuring input
Measuring range

0 ... ±10 V

Accuracy

error < 0.003 % rg.

3

typ.,

< 0.02 % rg. guar.

Input impedance

1 MW

Connection

4 mm (0.16 in) banana sockets

DC current measuring input
Measuring range

0 ... ±1 mA, 0 ... ±20 mA

Accuracy

error < 0.003 % rg.

3

typ.,

< 0.02 % rg. guar.

Input impedance

15 W

Connection

4 mm (0.16 in) banana socket

Analog AC+DC measuring inputs

4

Type

AC + DC analog voltage inputs
(current measurement with external
current clamps or shunt resistors)

Number

10

Nominal input ranges (RMS values)

100 mV, 1 V, 10 V, 100 V, 600 V

Amplitude accuracy

error < 0.06 % typ., < 0.15 % guar.

Bandwidth

DC ... 10 kHz

Sampling frequency

28.44 kHz, 9.48 kHz, 3.16 kHz

Input impedance

500 kW // 50 pF

Transient input buffer at 28 kHz

3.5 s for 10 input channels / 35 s for 1
input channel

Transient input buffer at 3 kHz

31 s for 10 input channels / 5 min. for 1
input channel

Transient trigger

threshold voltage, power quality trigger:
sag, swell, harmonic, frequency,
frequency change, notch

Measurement functions

I (AC + DC), V (AC + DC), phase,
frequency, power, harmonics,
transient recording, event recording,
trend recording

Input overload indication

yes

Input protection

yes

Max. input voltage

CAT IV / 150 V, CAT III / 300V, CAT II /
600 V (850 Vpk)

Galvanic isolation

5 groups (2+2+2+2+2)

Connection

4 mm (0.16 in) banana sockets
(combined with binary inputs)

Time Synchronization
Timing accuracy

IRIG-B synchronization with CMIRIG-B
GPS synchronization with CMGPS

error < 1 µs typ., < 5 µs guar.
error < 1 µs typ., < 5 µs guar.

To external voltage

Reference signal on binary input 10:
10 ... 300 V / 40 ... 70 Hz

With the unique PermaSync functionality (supported by TU 2.30 or higher), analog
and Sampled Values outputs stay permanently in sync with the internal CMC time
reference. In combination with the optional CMIRIG-B interface box, PermaSync
additionally allows the continuous synchronization of the output quantities with
an external IRIG-B time protocol or an external PPS signal. With CMIRIG-B it is also
possible to transmit the internal PPS signal of the CMC to the device under test (e.g.
PMUs or IEDs stimulated with a synchronized Sampled Values data stream).

1

For directly testing relays with low level inputs by simulating signals from non

conventional CTs and VTs with low level interfaces and for controlling external

amplifier units

2

THD+N: Values at 50/60 Hz, 20 kHz measurement bandwidth, nominal value, and

nominal load

3

rd. = reading, rg. = range

4

Up to three inputs can be used for measuring RMS values, frequency, and phase

angle without the EnerLyzer software license. Full functionality requires EnerLyzer

software license