Atec Quadtech-GUARDIAN-5000 User Manual
5000 safetyanalyzer, Guardian
5000 SafetyAnalyzer
USES:
• Production and Compliance Testing of
Appliances, Instruments and
Information Technology Equipment in
Accordance with UL, CSA, IEC, TUV
and Other Standards such as
EN60335, EN60950, EN61010, CSA
C22.2 No. 1010.1, UL3111 and
UL1950
• Transformer Electrical Safety Testing
• Electric Motor Safety Testing
• Power Supply Safety Testing
• Verification of the Ground Connection
on Products with a Three Prong Power
Cord
FEATURES:
• Programmable output voltage to 5KV
AC and 6KV DC
• Ground Bond Testing to 30A AC with
Adjustable Limit
• Multi-point Scanner Option for High
Voltage Switching of up to 64 Points
• Programmable Ramp and Test Times
• Storage of 50 Tests Setups with 15
Steps per Setup
• Trip Current Programmable to 40mA
AC and 20mA DC
• Front Panel Lockout via Password
• Standard IEEE and Remote Control
Interfaces
• Optional RS232 Interface
• LabVIEW® Intrument Driver
Software
• Insulation Resistance Measurements
from 10M
Ω
Ω to 10GΩ
Ω 1ohm Limit
Introduction
The Guardian 5000 is four instruments in one providing AC Hipot, DC Hipot,
Insulation Resistance, and Ground Bond measurements from a single test connector
in one versatile instrument. This provides a cost effective solution to electrical safe-
ty compliance testing with maximum flexibility for present and future requirements.
Performing all of the electrical safety tests in one box can reduce test time, increase
productivity and reduce the number of boxes requiring calibration.
Description
The Guardian 5000 AC Hipot
Tester performs AC dielectric testing (hipot) over the
voltage range from 100V to 5000VAC RMS. The maximum leakage current of 40mA
RMS makes the Guardian 5000 ideal for testing devices with high leakage currents
such as power supplies which have large filter or “Y” capacitors for noise reduction.
Both the actual test voltage and total current are displayed during testing.
The Guardian 5000 DC Hipot
Tester voltage range for DC dielectric testing is 100V
to 6000VDC with a resolution of 10V. The maximum current leakage is 20mA
which allows quick charging of capacitive devices. A quick discharge of the device
when the measurement is complete, also minimizes test times.
The Guardian 5000 Insulation Resistance measurements are similar to a DC hipot
but rather than displaying leakage current, resistance is calculated and displayed. The
insulation resistance can be measured over the range of 10M
Ω
to 10G
Ω
with test
voltages from 50VDC to 1000VDC in 1V steps.
The Guardian 5000 Ground Bond provides up to 30A AC for ground bond testing.
The test current can be programmed from 1A to 30A in 0.1A steps. Hi current limit,
test time, frequency and open circuit no load voltage can all be programmed. The
offset feature can automatically compensate for any lead resistance.
1-800-253-1230
Fax 1-978-461-4295
Intl. 1-978-461-2100
For more information about special
purchase, rent & lease options, call
For more detailed
specifications, visit
www.quadtech.com
AC/DC/IR/GC Electrical Safety Analyzer
Guardian