Main Screen / Front Panel Details
Test Status Indicator Bar
The test status indicator bar includes three sets of test
indicators to aid during testing. On the left side are the user
selected
Range, Low Pass Filter
and
High Pass Filter
settings.
On the right side are indicator lights: a
uto-range
shows if the
auto-range feature is enabled and
auto-ranging
lights when
the device is actively auto-ranging. The
Voltage
and
PD Meter
O/R
light when the meter reading is beyond the scale of the
meters. The light in the middle is the pass/fail indicator for PD
levels that are over the user defined limit and shows in red
during a “failed” test.
Function Menu Bar
The six buttons along side (the function menu bar) are used to
enter the parameters and select settings for the unit and the
test. The function menu bar along with the up/down buttons
supply the user with control and the ability to set/alter the
test setup.
Main Screen
The test results are displayed on the main screen and two
graphical modes are available to view the test results: meter
mode and scope mode. The threshold shown graphically (in
color) together with different meter modes makes monitoring
of the test results simple.
When operating in the
meter mode,
you have opted to only
view the magnitudes of the partial discharge. In meter mode,
two meters are displayed: one for partial discharge and one
for voltage. You can select two different types of meters: a
traditional analogue meter display or a bar-type digital
display. In either mode you select the update rate as either
fast or slow.
The
scope mode
display looks just like an oscilloscope. You
select the type of oscilloscope display as an ellipse, straight
line or a sine wave. Gating features are also available in scope
mode.
Scope mode gives
you the ability to
see more than just
the magnitude of
the PD and allows
a
knowledgeable
user to perform
diagnosis of the
source of the PD.
You
can
also
choose
whether
there’s a single or
double gate, the
start point, and
the width to gate
out
undesirable
interferences.