DC VOLTAGE

ACCURACY 23°C ± 5°C

±(ppm of rdg. + ppm of range)

INPUT

RANGE

RESOLUTION

90 DAY

1 YEAR

RESISTANCE

100.0000 mV

0.1 µV

40 + 35

50 + 35

> 10 G

1.000000

V

1.0 µV

25 + 7

30 + 7

> 10 G

10.00000

V

10 µV

20 + 5

30 + 5

> 10 G

100.0000

V

100 µV

30 + 6

45 + 6

10 M

Ω ±1%

1000.000

V

1 mV

35 + 6

45 + 6

10 M

Ω ±1%

RESISTANCE

ACCURACY 23°C ± 5°C

±(ppm of rdg. + ppm of range)

TEST

RANGE

RESOLUTION

90 DAY

1 YEAR

CURRENT

100.0000

100 µ

80 + 40

100 + 40

1 mA

1.000000 k

1 m

80 + 10

100 + 10

1 mA

10.00000 k

10 m

80 + 10

100 + 10

100 µA

100.0000 k

100 m

80 + 10

100 + 10

10 µA

1.000000 M

1

80 + 10

100 + 10

10 µA

10.00000 M

10

200 + 10

400 + 10

700 nA

100.0000 M

100

1500 + 30

1500 + 30

700 nA

DC CURRENT

ACCURACY 23°C ± 5°C

±(ppm of rdg. + ppm of range)

BURDEN

RANGE

RESOLUTION

90 DAY

1 YEAR

VOLTAGE

10.00000 mA

10 nA

300 + 80

500 + 80

< 0.15 V

100.0000 mA

100 nA

300 + 800

500 + 800

< 0.03 V

1.000000

A

1 µA

500 + 80

800 + 80

< 0.3 V

3.00000

A

10 µA

1200 + 40

1200 + 40

< 1 V

CONTINUITY 2W

ACCURACY 23°C ± 5°C

±(ppm of rdg. + ppm of range)

TEST

RANGE

RESOLUTION

90 DAY

1 YEAR

CURRENT

1 k

100 m

100 + 100

120 + 100

1 mA

DIODE TEST

ACCURACY 23°C ± 5°C

±(ppm of rdg. + ppm of range)

TEST

RANGE

RESOLUTION

90 DAY

1 YEAR

CURRENT

3.00000

V

10 µV

30 + 7

40 + 7

1 mA

10.00000

V

10 µV

30 + 7

40 + 7

100 µA

10.00000

V

10 µV

30 + 7

40 + 7

10 µA

DC OPERATING CHARACTERISTICS

RMS NOISE

DIGITS

READINGS/s

PLCs

10V RANGE

NMRR

CMRR

6

1

2

5

10

< 1.5 µV

60 dB

140 dB

6

1

2

50

1

< 4 µV

60 dB

140 dB

5

1

2

500

0.1

< 22 µV

80 dB

4

1

2

2000

0.01

< 150 µV

80 dB

DC SYSTEM SPEEDS

2, 6

RANGE CHANGE

3

: 50 / s.

FUNCTION CHANGE

3

: 45 / s.

AUTORANGE TIME

3, 10

: <30 ms.

ASCII READINGS TO RS-232 (19.2K BAUD): 55 / s.

MAX. INTERNAL TRIGGER RATE: 2000 / s.

MAX. EXTERNAL TRIGGER RATE: 400 / s.

TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS

FREQUENCY

ACCURACY (I Year) , 23°C ±5 °C

RANGE

RESOLUTION

RANGE

±(% of reading + % of range)

100 mV to 750 V

0.1 µV to 1 mV

3 Hz–10 Hz

0.35 + 0.03

10 Hz–20 kHz

0.06 + 0.03

20 kHz–50 kHz

0.12 + 0.05

50 kHz–100 kHz

0.60 + 0.08

100 kHz–300 kHz

4 + 0.5

AC OPERATING CHARACTERISTICS

FUNCTION

DIGITS

READINGS/s

RATE

BANDWIDTH

ACV (all ranges), and

6

1

2

2s/reading

SLOW

3 Hz–300 kHz

ACI (all ranges)

6

1

2

1.4

MED

30 Hz–300 kHz

6

1

2

4.8

MED

30 Hz–300 kHz

6

1

2

2.2

FAST

300 Hz–300 kHz

6

1

2

35

FAST

300 Hz–300 kHz

FREQUENCY AND PERIOD CHARACTERISTICS

RESOLUTION

ACCURACY

ACV

FREQUENCY

PERIOD

GATE

±(ppm of

90 DAY/1 YEAR

RANGE

RANGE

RANGE

TIME

reading)

±(% of reading)

100 mV

3 Hz

333 ms

1 s (SLOW)

0.333

0.01

to

to

to

0.1 s (MED)

3.33

0.01

750 V

500 kHz

2 µs

10 ms (FAST)

33.3

0.01

FREQUENCY NOTES

1. Specifications are for squarewave inputs >10% of ACV range, except 100mV range. On 100mV range frequency must

be >10Hz if voltage is <20mV.

2. 20% overrange on all ranges except 750V range.

TEMPERATURE CHARACTERISTICS

THERMOCOUPLE

90 DAY/1 YEAR (23°C ± 5°C)

ACCURACY

Relative to

USING

TYPE

RANGE

RESOLUTION

Reference Junction

2001-TCSCAN

5

J

–200 to +

760°C

0.001°C

±0.5°C

±0.65°C

K

–200 to + 1372°C

0.001°C

±0.5°C

±0.70°C

T

–200 to +

400°C

0.001°C

±0.5°C

±0.68°C

GENERAL SPECIFICATIONS

POWER SUPPLY: 100V / 120V / 220V / 240V ±10%.

LINE FREQUENCY: 45Hz to 66Hz, automatically sensed at power-up.

POWER CONSUMPTION: 22VA.

OPERATING ENVIRONMENT: Specified for 0°C to 50°C. Specified to 80% R.H. at 35°C.

STORAGE ENVIRONMENT: –40°C to 70°C.

WARRANTY: 3 years.

SAFETY: Conforms with European Union Directive 73/23/EEC, EN 610110-1,

UL 3111-1.

EMC: Conforms with European Union Directive 89/336/EEC, EN 55011, EN 50082-1, EN 61000-3-

2, EN 61000-3-3, FCC part 15 class B.

WARMUP: 1 hour to rated accuracy.

DIMENSIONS: Rack Mounting: 89mm high

× 213mm wide × 370mm deep (3

1

2

in

× 8

3

8

in

×

14

9

16

in).

Bench Configuration (with handle and feet): 104mm high

× 238mm wide × 370mm deep

(4

1

8

in

× 9

3

8

in

× 14

9

16

in).

NET WEIGHT: 2.9kg (6.3 lbs).

SHIPPING WEIGHT: 5kg (11 lbs).

VOLT HERTZ PRODUCT:

≤8 × 10

7

V·Hz.

ACCESSORIES SUPPLIED: Model 1751 Safety Test Leads, User Manual.

2015, 2015-P,

6

1

2

-Digit THD Multimeters

2016, 2016-P

6

1

2

-Digit Audio Analyzing Multimeters