Ifr-1900 csa, Wireless - radio test sets ifr-1900 csa – Atec IFR-1900CSA User Manual
Page 3

Frequency Range
10 MHz to 2010 MHz
Accuracy
Same as Master Oscillator
Resolution
1 Hz (fc <20 MHz)
10 Hz (fc >20 MHz)
Minimum Level
-60 dBm (ANT connector)
Frequency Digital Meter Range
0 Hz to ±150 kHz
Bar Graph Meter Range
0 to ±100 kHz (in 4 decade ranges)
Accuracy
Same as Master Oscillator ± LSD
Resolution
1 Hz (±1 Hz to ±10 kHz)
10 Hz (>±10 kHz to ±150 kHz)
Minimum Level
-60 dBm (ANT Input Port)
Frequency Range
100 MHz to 2010 MHz
Input Level
0.05 mW to 50 W RMS (<900 MHz, 125
sequence, 4 decade)
0.05 mW to 10 W RMS (>900 MHz, 125
sequence, 4 decade)
Resolution
1 %
Accuracy
±6% (>5 W and <50 W, at Typical Operational
Ambient Temperature)
LOW LEVEL POWER METER
Frequency Range
Same as standard RF power meter
Input level
-40 dBm to -10 dBm
Accuracy
12% typical
Frequency Range
10 MHz to 2010 MHz
Sensitivity
<-80 for 10 dB SINAD
(1 kHz rate, 6 kHz Dev, FM 2, ANT Input Port)
Demodulation Output Level
(FM): 5 Vp-p ±15 % (at full scale into 600
Ω
)
(PM): 40 mVRMS ±15 % (5 Rad, into 600
Ω
)
(AM): 1 VRMS ±15 % (80 % modulation, into
600
Ω
)
(SSB): 1.15 VRMS ±15 % (Beat tone, into 600
Ω
)
Receive IF Output Signal Frequency
88 MHz to 90 MHz
IF Bandwidth
8.5 MHz
Deviation Range
±100 Hz to ±100 kHz
Resolution
100 Hz (20 kHz ranges)
1 kHz (>20 kHz ranges)
Accuracy
±5 % ±2 counts + source residual FM
(300 kHz IF, <15 kHz rate)
Modulation Rate
100 Hz to 40 kHz
Carrier Range
100 MHz to 2010 MHz
Minimum Carrier Level
-60 dBm (ANT Input Port)
Deviation Range
0 Rad to 10 Rad (Peak)
Resolution
0.01 Rad (deviation
≤
5 Rad)
0.1 Rad (deviation >5 Rad)
Carrier Range
100 MHz to 2010 MHz
Minimum Carrier Level
-60 dBm (ANT Input Port)
Modulation Range
1 % to 90 %
Resolution
1 %
Accuracy
±5 % of full scale ±1 count + source residual AM
(30 % to 90 %)
Carrier Range
100 MHz to 2010 MHz
Minimum Carrier Level
-60 dBm (ANT Input Port)
Distortion Range
0.1 % to 20 %
Resolution
0.1 %
Accuracy
±0.5 % distortion ±1 count (1 % to 10 %)
±2 % distortion ±1 count (>10 %)
Signal Frequency
700 Hz to 1.4 kHz
Signal Level
0.1 VRMS to 30 VRMS (SINAD/BER input)
Input Range
NT 400 Channels
Cellular 800 MHz Channels
PCS 1900 MHz Channels
Minimum Carrier Level
-60 dBm (ANT connector)
EVM Range
0 to 100 %
EVM Resolution
0.01 %
Meter Residual EVM
<2 % indication
Accuracy
± 3.0% indication, ± 1 LSD + meter residual EVM
Range
3 dB to 30 dB
Resolution
0.1 dB
Accuracy
±1 dB ±1 count (at 12 dB SINAD)
Signal Frequency
700 Hz to 1.4 kHz
Signal Level
0.1 VRMS to 30 VRMS (SINAD/BER input)
VOLTMETER (DC/AC)
Ranges
200 mV to 200 V (full scale, decade sequence,
150
Ω
, 600
Ω,
1 M
Ω
selectable
)
Maximum Input
1000 VDC
500 VAC
Resolution
3.5 digit (maximum resolution 0.1 mV on 200 mV
range)
Accuracy
±5% of full scale ± 1 count (AC, where ACV*kHz
< 140)
±1% of full scale ± 1 count (DC)
Frequency
DC, 50 Hz to 20 kHz
Input Impedance
1 M
Ω
(±5%)
150
Ω
(±5%)
600
Ω
(±5%)
OHMMETER
Ranges
200
Ω
to 20 M
Ω
(full scale, decade sequence)
Resolution
3.5 digit (maximum resolution 0.1 on 200 range)
Accuracy
±5% or 0.1 (±1 count)
CURRENT METER (DC/AC)
Ranges
20 mA to 2 A (full scale, decade sequence, 20 A
maximum when using external shunt)
Resolution
3.5 digit (maximum resolution 0.01 mA on 20 mA
range)
Accuracy
±5 % or 0.1 mA ±1 count
Vertical BW
1 MHz (-3 dB)
Input Ranges
1 mV / Div to 50 V / Div (1-2-5 sequence, 8
divisions)
Max Input Vertical
200 V Peak
Accuracy Vertical
±5% of full scale
±10% of full scale (1 mV and 2 mV ranges)
Resolution Vertical
Full Scale / 256
Coupling Vertical
AC, DC, GND
Horizontal Sweep Rate
1 µSec / Div to 100 msec / Div (1-2-5 sequence,
10 divisions)
Accuracy Horizontal
±1 % of Full Scale
Resolution Horizontal
Full Scale / 400
External Input Impedance Horizontal
1 M
Ω
(shunted by 27 pF nominal)
Internal Signal Routing
425 kHz IF
Demod Audio
Function Generator
SINAD/BER
RF Power
EXT MOD
Range
10 MHz to 2010 MHz
Frequency Span Range
1 kHz/Div to 100 MHz/Div plus Zero Scan
(10 divisions in a 1-2-5 sequence)
Accuracy
±5% of Span Width
Reference Accuracy
See Master Oscillator
Display
Log, 10 dB/Div and 2 dB/Div
Vertical Resolution
Full Scale/256
Displayed Range (Dynamic)
60 dB (0 dB Attenuation, Span <1 MHz/Div)
Bandwidth Switching Error
<2 dB (5 kHz/Div thru 1 MHz/Div)
<3 dB (<5 kHz/Div or >1 MHz/Div)
Spectrum Analyzer
Oscilloscope
Digital Multimeter
SINAD Meter
Error Vector Magnitude (EVM) Meter
Distortion Meter
AM Modulation Meter
PM Deviation Meter
FM Deviation Meter
Receiver
RF Power Meter
RF Frequency Error Meter
RF Counter
IFR-1900 CSA
W
ireless - Radio test Sets
IFR-1900 CSA
3