** Note: all specifications for tests on twisted-pair cabling apply to 100

Ω cable. Contact Fluke Networks for information on measurement

performance for cable with a different impedance.

Length specs do not include the uncertainty of the cable’s NVP Value

Above specifications are subject to change without notice.

Environmental Specifications

Operating temperature

32° F to 104° F (0° C to 45° C)

Storage temperature

-4° F to +140° F (-20° C to +60° C)

Operating relative humidity

32° F to 95° F (0° C to 35° C): 0 % to 90 %

(% RH without condensation)

95° F to 113° F (35° C to 45° C): 0 % to 70 %

Vibration

Random, 2 g, 5 Hz -500 Hz

Shock

1 m drop test with and without module and adapter attached

Safety

CSA C22.2 No. 1010.1: 1992

EN 61010-1 1st Edition + Amendments 1, 2

Pollution degree

Pollution Degree 2 as defined in IEC 60664 and applied to IEC 60950 ‘Safety of Information

Technology Equipment, 1999’

Altitude

Operating: 4000 m; Storage: 12000 m

EMC

EN 61326-1

Performance Specifications**

Baseline accuracy

The baseline accuracy of all DTX Series testers meets or exceeds the requirements of Level IV as in

draft 2nd Edition of IEC 61935-1 over the frequency range supported by the tester.

Category 6/Class E link test modes

DTX CableAnalyzer far exceeds Level III requirements of TIA/EIA-568-B.2-2 and IEC 61935-1.

(and lower link categories)

Class F link test modes

DTX-1800 is compliant with Level IV requirements as in the draft 2nd edition of IEC 61935-1.

Length for Twisted Pair Cabling

Parameter

Single-ended Test

Dual-ended Test (Main and Remote)

Range

800 m (2600 ft)

150 m (490 ft)

Resolution

0.1 m or 1 ft

0.1 m or 1 ft

Accuracy

± (.3 m + 2%); for lengths 0 to 150 m

± (.3 m + 2%)

± (.3 m + 4%); for lengths 151 to 800 m

Propagation Delay

Parameter

Single-ended Test

Dual-ended Test (Main and Remote)

Range

4000 ns

750 ns

Resolution

1 ns

1 ns

Accuracy

± (2 ns + 2%); 0 to 750 ns

± (2 ns + 2%)

± (2 ns + 4%); 751 to 4000 ns

Delay Skew

Parameter

Twisted Pair Cabling

Range

0 ns to 100 ns

Resolution

1 ns

Accuracy

± 10 ns

DC Loop Resistance Test

Parameter

Twisted Pair Cabling

Range

O

Ω to 53 Ω

Resolution

0.1

Accuracy

± (1

Ω + 1%)

Overload recovery time

Less than 10 minutes to rated accuracy following over-voltage. Referencing is required after

repeated or prolonged over-voltage.

Fluke Corporation
P.O. Box 777, Everett, WA USA 98206-0777

Fluke Networks operates in more than 50 countries
worldwide. To find your local office contact details,
go to

www.flukenetworks.com/contact

.

©2004 Fluke Corporation. All rights reserved.
Printed in U.S.A. 6/2004 2131195 B-ENG-N Rev B

N E T W O R K

S U P E R

V I S I O N