CertiFiber Pro Optical Loss Test Set Module Specifications
Power Meter Specifications
Input Connector
Interchangeable connector adapter (LC standard, SC, ST and FC optional)
Detector Type
InGaAs
Wavelengths
850 nm, 1300 nm, 1310 nm, 1550 nm
Power Measurement Range
0 dBm to -65 dBm (850 nm) 0 dBm to -70 dBm (all other wavelengths)
Power Measurement Uncertainty
1,2
< +/- 5% +/- 32 pW
Measurement Linearity
3
< ± 0.1 dB
Re-calibration Period
1 year
1. +/- 100 pW at 850 nm
2. Under the following conditions: Power level 100 μW (-10 dBm), continuous wave (CW) for absolute power at 850 nm and 1310 nm. Divergent beam, NA = 0.20 for
50/125 μm and NA = 0.14 for 9/125 μm. Ambient temperature 23° ± 1°C. SC/UPC connector with ceramic ferrule. After a 5 minute warm-up. Traceable to NIST.
3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23° ± 1°C. After a 5 minute warm-up.
Loss/Length Specifications
Specification
CertiFiber Pro Multimode Modules
CertiFiber Pro Singlemode Modules
Testing Speeds (excludes referencing times)
Smart Remote mode: < 3 seconds (2 wavelengths, one direction, auto wavelength detection)
Far End Source mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection)
Loopback mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection)
Input/Output Connectors
Interchangeable connector adapter (LC standard, SC, ST and FC optional)
Launch Condition
1,2
Encircled Flux compliant to TIA-526-14-B, ISO/IEC 14763-3 and IEC 61280-4-1
Fiber Types Tested
50/125 µm or 62.5/125 µm
singlemode
Source Type and Wavelengths
LED source 850 nm ± 30 nm 1300 nm ± 20 nm
Fabry-Perot laser diode 1310 nm ± 20 nm 1550 nm ± 30 nm
Maximum Length Measurement
12 km
130 km
Length Measurement Accuracy
± 1.5 m plus ± 1 % of length
Output Power (Nominal)
≥ -24 dBm with EF-TRC
≥ -4 dBm
Output Power Stability
3
± 0.05 dB over 8 hours ± 0.03 dB over 15 minutes
± 0.1 dB over 8 hours ± 0.08 dB over 15 minutes
1. At the output of the EF-TRC
2. Variations between EF measurement equipment may occur but EF compliance can be expected with a 95% confidence factor
3. Relative to power level after 15 minute warm up – at constant temperature