beautypg.com

Ppm-350b, Ppm-352b-eg-er- xx, Pon power meter – Atec Exfo-PPM-350B-EG User Manual

Page 4: Ordering information

background image

SPPPM-350B.8AN

© 2007 EXFO Electro-Optical Engineering Inc. All rights reserved.

Printed in Canada 07/11

EXFO Corporate Headquarters

> 400 Godin Avenue, Quebec City (Quebec) G1M 2K2 CANADA Tel.: 1 418 683-0211 Fax: 1 418 683-2170 [email protected]

Toll-free: 1 800 663-3936 (USA and Canada)

www.EXFO.com

EXFO is certified ISO 9001 and attests to the quality of these products. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation. EXFO has made every effort to ensure that the information contained in this specification sheet is accurate. However, we accept no responsibility for any errors or omissions, and we reserve the right to modify
design, characteristics and products at any time without obligation. Units of measurement in this document conform to SI standards and practices. In addition, all of EXFO’s manufactured products are compliant with the European Union’s WEEE directive. For more information, please
visit

www.EXFO.com/recycle

. Contact EXFO for prices and availability or to obtain the phone number of your local EXFO distributor.

For the most recent version of this spec sheet, please go to the EXFO website at

http://www.EXFO.com/specs

In case of discrepancy, the Web version takes precedence over any printed literature.

EXFO America

3701 Plano Parkway, Suite 160

Plano, TX 75075 USA

Tel.: 1 800 663-3936

Fax: 1 972 836-0164

EXFO Europe

Omega Enterprise Park, Electron Way

Chandlers Ford, Hampshire S053 4SE ENGLAND

Tel.: +44 2380 246810

Fax: +44 2380 246801

EXFO Asia

151 Chin Swee Road, #03-29 Manhattan House

SINGAPORE 169876

Tel.: +65 6333 8241

Fax: +65 6333 8242

EXFO China

No. 88 Fuhua, First Road, Central Tower, Room 801

Shenzhen 518048, CHINA

Tel.: +86 (755) 8203 2300

Fax: +86 (755) 8203 2306

Futian District
Beijing New Century Hotel Office Tower, Room 1754-1755

Beijing 100044 P. R. CHINA

Tel.: +86 (10) 6849 2738

Fax: +86 (10) 6849 2662

No. 6 Southern Capital Gym Road

PPM-350B

PON Power Meter

Rugged Handheld Solutions

OTDRs
OLTSs
Power meters
Light sources
Talk sets

OPTICAL

ADSL/ADSL2+,
SHDSL, VDSL test sets
VoIP and IPTV test sets
Ethernet test sets
POTS test sets

COPPER ACCESS

Platform-Based Solutions

OTDRs
OLTSs
ORL meters
Variable attenuators

OPTICAL FIBER

OSAs
PMD analyzers
Chromatic
dispersion analyzer

DWDM TEST SYSTEMS

Next-generation SONET/SDH and OTN testers
SONET/DSn (DS0 to OC-192) testers
SDH/PDH (64 kbit/s to STM-64) testers
T1/T3, E1 testers
10/100 Mbit/s and Gigabit Ethernet testers
Fibre Channel testers
10 Gigabit Ethernet testers

TRANSPORT AND DATACOM

ORDERING INFORMATION

PPM-352B-EG-ER-

XX

SPECIFICATIONS

a

PPM-352B-EG-ER

BPON

EPON/GPON

Power measurement range – pass zone

1310 nm

10 to –40

for continuous data stream (dBm)

1490 nm

12 to –40

1550 nm

25

b

to –40

Burst mode measurement capability:

CO to ONT

Burst mode measurement range

b

(dBm)

1310 nm

10 to –33

10 to –29

ORL

c

(dB)

1550 nm

55

Pass through insertion loss

b

(dB)

1.5

Spectral passband (nm)

1310 nm

1260 to 1360

1490 nm

1480 to 1500

1550 nm

1539 to 1565

Power uncertainty at calibrated

0.5

wavelengths

b, d

(dB)

Refresh rate of display (Hz)

2.5

Calibrated wavelengths (nm)

1310, 1490, 1550

Threshold sets

10 configurable threshold sets with threshold naming

Autonomy

b

(hours)

> 30

Number of ports

2

Warranty and recommended

1

calibration interval (year)

User guide, three AA batteries, wrist strap, PC threshold-transfer software, RS-232 cable.

STANDARD ACCESSORIES

Size (H x W x D)

185 mm x 100 mm x 55 mm

(7

1

/

4

in x 4 in x 2

1

/

8

in)

Weight

0.4 kg

(0.9 lb)

Temperature

operating –10 °C to 50 °C

(14 °F to 122 °F)

storage

–40 °C to 70 °C

(–40 °F to 158 °F)

Relative humidity

0 % to 95 % non-condensing

GENERAL SPECIFICATIONS

Model
PPM-35

2

B-

EG-ER

= PON power meter, two ports, BPON,

extended range EPON, GPON

Connector

e

EI-EUI-

28

= UPC/DIN 47256

EI-EUI-

76

= UPC/HMS-10/AG

EI-EUI-

89

= UPC/FC narrow key

EI-EUI-

90

= UPC/ST

EI-EUI-

91

= UPC/SC

EI-EUI-

95

= UPC/E-2000

EA-EUI-

28

= APC/DIN 47256

EA-EUI-

89

= APC/FC narrow key

EA-EUI-

91

= APC/SC

EA-EUI-

95

= APC/E-2000

Example: PPM-352B-EG-ER-EA-EUI-91

Notes

a. At room temperature.

b. Typical.

c. For APC connectors. Typically > 35 dB for UPC connectors.

d. Around –7 dBm, CW.

e. Same connectors for both ports.