beautypg.com

Rf power meter (standard) – Atec JDSU-JD748A User Manual

Page 3

background image

3

JD748A CELLADVISOR SIgnAL AnALYZER

Spurious

Inherent residual response
Input terminated, 0 dB attenuation, preamplifier off, RBW at 10 kHz
20 MHz to 3 GHz

–90 dBm (nominal)

>3 GHz to 4 GHz

–85 dBm (nominal)

Exceptions

<–80 dBm @ 311.94 MHz
<–84 dBm @ 415.92 MHz
<–85 dBm @ 519.90, 1599.00, and 2497.80 MHz

Input related spurious <–70 dBc (nominal)

Dynamic Range

2/3 (TOI-DANL) in 1 Hz RBW

>95 dB

Sweep Time

Range

80 ms to 1000 s
24 µs to 200 s


Span = 0 Hz (zero span)

Sweep mode

Continuous, single

Gated Sweep

Trigger source

External, video, and GPS

Gate length

1 µs to 100 ms

Gate delay

0 to 100 ms

Trigger

Trigger source

Free run, video, external

Trigger delay
Range
Resolution


0 to 200 s
6 µs

Measurements

Channel power
Occupied bandwidth
Spectrum emission mask
Adjacent channel power
Spurious emissions
Field strength
AM/FM audio demodulation
Route map
PIM detect
Dual spectrum

RF Power Meter (Standard)

General Parameters

Display range

–100 to +100 dBm

Offset range

0 to 60 dB

Resolution

0.01 dB or 0.1xW (x = m, u, p)

Internal RF Power Sensor

Frequency range

10 MHz to 4 GHz

Span

100 kHz to 100 MHz

Dynamic range

–120 to +20 dBm

Maximum power

+20 dBm

Accuracy

Same as spectrum analyzer

External RF Power Sensors

Directional Power Sensor

JD731B

Frequency range

300 MHz to 3.8 GHz

Dynamic range

0.15 to 150 W (average)
4 to 400 W (peak)

Connector type

Type-N female on both ends

Measurement type

Forward/reverse average power,
forward peak power, VSWR

Accuracy

±(4% of reading + 0.05 W)

1,2

Directional Power Sensor
Frequency range

JD733A
150 MHz to 3.5 GHz

Dynamic range

0.1 to 50 W (average)
0.1 to 50 W (peak)

Connector type

Type-N female on both ends

Measurement type

Forward/reverse average power,
forward peak power, VSWR

Accuracy

±(4% of reading + 0.05 W)

1,2

Terminating Power Sensor

JD732B

Frequency range

20 MHz to 3.8 GHz

Dynamic range

–30 to +20 dBm

Connector type

Type-N male

Measurement type

Average

Accuracy

±7%

1

Terminating Power Sensor

JD734B

Frequency range

20 MHz to 3.8 GHz

Dynamic range

–30 to +20 dBm

Connector type

Type-N male

Measurement type

Peak

Accuracy

±7%

1

Terminating Power Sensor

JD736B

Frequency range

20 MHz to 3.8 GHz

Dynamic range

–30 to +20 dBm

Connector type

Type-N male

Measurement type

Average and Peak

Accuracy

±7%

1

1. CW condition at 25°C ±10°C.

2. Forward power.