beautypg.com

Atec Agilent-8568b-8566b User Manual

Page 5

background image

5

CISPR EMI Amplitude

Parameter

Specification

Measurement range

+137 dBµV to noise floor (refer to amplitude sensitivity table)

Amplitude accuracy8
(+10° to +30°C, 0 – 20 dB RFP atten.)

±2.0 dB for 10 dB SA atten.
± 2.5 dB for 0 dB SA atten.

Calibration10

Automatic calibration for absolute amplitude with comb
generator (RFP)

Linearity check

3 dB input attenuator test for system overload and distortion
(RFP)

Impulsive signal tolerance

Input #1 withstands LISN transients (RFP)
Input #2 withstands CISPR calib. pulses (RFP)

Detection9

Peak and average (SA), quasi-peak (QPA)

Audio output

Speaker with volume control (QPA)

CISPR EMI Amplitude sensitivity

EMI receiver

Specification

Frequency band

Resolution bandwidth (6 dB)

Average noise6
(RFP/SA atten. = 0 dB)

Quasi-peak noise
(RFP/SA atten. = 0 dB)

10 – 150 kHz (band A)

200 Hz

-12 dBµV4
-17 dBµV typical

-15 dBµV (nominal)

150 kHz – 1 MHz (band B)

9 kHz

+4 dBµV
-1 dBµV typical

0 dBµV (nominal)

1 – 30 MHz (band B)

9 kHz

-13 dBµV
-16 dBµV typical

-11 dBµV (nominal)

30 – 1000 MHz (band C, D)

120 kHz

-1 dBµV
-6 dBµV typical

0 dBµV (nominal)