Spectrum Master™ MS2712E and MS2713E Specifications

Power Meter (Option 0029)

Frequency

Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band

Amplitude

Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale

Average

Acquisition Fast/Med/Slow, # of Running Averages

Limits

Limit On/Off, Limit Upper/Lower

Frequency Range

10 MHz to 4 GHz (MS2712E), 10 MHz to 6 GHz (MS2713E)

Span

1 kHz to 100 MHz

Display Range

-140 dBm to +30 dBm, ≤40 dB span

Measurement Range

-120 dBm to +26 dBm

Offset Range

0 to +100 dB

VSWR

2:1 typical

Maximum Power

+26 dBm without attenuator

Accuracy

Same as Spectrum Analyzer

Application Options

Impedance (50 , 75 , Other)

High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))

Amplitude

Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale

Average

# of Running Averages, Max Hold

Zero/Cal

Zero On/Off, Cal Factor (Center Frequency, Signal Standard)

Limits

Limit On/Off, Limit Upper/Lower

Power Sensor Model

PSN50

MA24104A

MA24106A

MA24108A

MA24118A

Description

High Accuracy

RF Power Sensor

Inline High

Power Sensor

High Accuracy

RF Power Sensor

Microwave USB

Power Sensor

Microwave USB

Power Sensor

Frequency Range

50 MHz to 6 GHz

600 MHz to 4 GHz

50 MHz to 6 GHz

10 MHz to 8 GHz

10 MHz to 18 GHz

Connector

Type N(m), 50

Type N(m), 50

Type N(m), 50

Type N(m), 50

Type N(m), 50

Dynamic Range

-30 to +20 dBm

(.001 to 100 mW)

+3 to +51.76 dBm

(2 mW to 150 W)

-40 to +23 dBm

(0.1 µW to 200 mW)

-40 to +20 dBm

(0.1 µW to 100 mW)

-40 to +20 dBm

(0.1 µW to 100 mW)

VBW

100 Hz

100 Hz

100 Hz

50 kHz

50 kHz

Measurand

True-RMS

True-RMS

True-RMS

True-RMS,

Slot Power,

Burst Average

Power

True-RMS,

Slot Power,

Burst Average

Power

Measurement Uncertainty

±0.16 dB

1

±0.17 dB

2

±0.16 dB

1

±0.18 dB

3

±0.18 dB

3

Datasheet

(for complete speciications)

11410-00414

11410-00483

11410-00424

11410-00504

11410-00504

Notes:

1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than -20 dBm with

zero mismatch errors.

2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.

Measurement results referenced to the input side of the sensor.

3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than-20 dBm with zero mismatch errors.