beautypg.com

Gps receiver option (option 0031) – Atec Anritsu-MS2721B User Manual

Page 4

background image

Page 4 of 20

Spectrum Master™ MS2721B Spectrum Analyzer Specifications

GPS Receiver Option (Option 0031)

(includes antenna)

Setup

On/Off, GPS Info

GPS Time/Location Indicator

Time, Latitude, Longitude and Altitude on display

Time, Latitude, Longitude and Altitude with trace storage

High Frequency Accuracy

Spectrum Analyzer, Interference Analyzer, Signal Analyzer

when GPS Antenna is connected

< ± 25 ppb with GPS On, 3 minutes after satellite lock in selected mode

GPS Lock – after antenna is disconnected

< ± 50 ppb for 3 days, 0 °C to 50 °C ambient temperature

Connector

BNC, female, reverse polarity

High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))

Amplitude

Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale

Average

# of Running Averages, Max Hold

Zero/Cal

Zero On/Off, Cal Factor (Center Frequency, Signal Standard)

Limits

Limit On/Off, Limit Upper/Lower

Power Sensor Model

PSN50

MA24104A

MA24106A

MA24108/18/26A

Description

High Accuracy

RF Power Sensor

Inline High

Power Sensor

High Accuracy

RF Power Sensor

Microwave

USB Power Sensor

Frequency Range

50 MHz to 6 GHz

600 MHz to 4 GHz

50 MHz to 6 GHz

10 MHz to 8 GHz

(MA24108A)

10 MHz to 18 GHz

(MA24118A)

10 MHz to 26 GHz

(MA24126A)

Connector

Type N(m), 50 Ω

Type N(m), 50 Ω

Type N(m), 50 Ω

Type N(m), 50 Ω

(MA24108/18A)

Type K(m), 50 Ω

(MA24126A)

Dynamic Range

–30 dBm to +20 dBm

(.001 mW to 100 mW)

+3 dBm to +51.76 dBm

(2 mW to 150 W)

–40 dBm to +23 dBm

(0.1 µW to 200 mW)

–40 dBm to +20 dBm

(0.1 µW to 100 mW)

VBW

100 Hz

100 Hz

100 Hz

50 kHz

Measurand

True-RMS

True-RMS

True-RMS

True-RMS, Slot Power,

Burst Average Power

Measurement Uncertainty

± 0.16 dB

1

± 0.17 dB

2

± 0.16 dB

1

± 0.18 dB

3

Datasheet

(for complete specifications)

11410-00414

11410-00483

11410-00424

11410-00504

Notes:

1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than –20 dBm with

zero mismatch errors.

2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.

Measurement results referenced to the input side of the sensor.

3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than –20 dBm with zero mismatch errors.