beautypg.com

General specifications – Atec Anritsu-MS2721B User Manual

Page 14

background image

Page 14 of 20

Spectrum Master™ MS2721B Spectrum Analyzer Specifications

General Specifications

All specifications and characteristics apply under the following conditions, unless otherwise stated: 1) After 5 minutes of warm-up time, where the
instrument is left in the ON state; 2) All specifications apply when using internal reference; 3) All specifications subject to change without notice;
4) Typical performance is the measured performance of an average unit; 5) Recommended calibration cycle is 12 months.

Maximum Continuous Input

(≥ 10 dB input attenuation) +30 dBm

Input Damage Level

≥ 10 dB input attenuation, > +43 dBm, ± 50 Vdc. Input protection relay opens at > 30 dBm
< 10 dB input attenuation, > +23 dBm, ± 50 Vdc. Limited dV/dt. Input protection relay
opens at approximately 10 to 23 dBm

ESD Damage Level

(≥ 10 dB input attenuation) > 10 kV

External Reference Frequencies

1 MHz, 1.2288 MHz, 1.544 MHz, 2.048 MHz, 2.4576 MHz, 4.8 MHz, 4.9152 MHz, 5 MHz,
9.8304 MHz, 10 MHz, 13 MHz and 19.6608 MHz at –10 dBm to +10 dBm

Setup Parameters

System

Status (Temperature, Battery Info, S/N, Firmware Ver, IP Address, Options Installed)
Self Test, Application Self Test
GPS (see Option 0031)

System Options

Name, Date and Time, Ethernet Configuration, Brightness, Volume
Language (English, French, German, Spanish, Chinese, Japanese, Korean, Italian, User defined)
Reset (Factory Defaults, Master Reset, Update Firmware)

File

Save, Recall, Delete, Directory Management

Save/Recall

Setups, Measurements, Screen Shots Jpeg (save only)

Delete

Selected File, All Measurements, All Mode Files, All Content

Directory Management

Sort Method (Name/Type/Date), Ascend/Descend, Internal/USB, Copy, Format USB

Internal Trace/Setup Memory

> 13,000 traces

External Trace/Setup Memory

Limited by size of USB Flash drive or Compact Flash module

Mode Switching

Auto-Stores/Recalls most recently used Setup Parameters in the Mode

Connectors

RF In

Type N, female, 50 Ω, Maximum Input +30 dBm, ± 50 VDC

GPS

BNC, female, reverse polarity

External Power

5.5 mm barrel connector, 12 to 15 VDC, < 5.0 Amps

LAN Connection

RJ48C, 10/100 Mbps, Connect to PC or LAN for Remote Access

Compact Flash Interface

Type I

USB Interface

Type A, Connect Flash Drive and Power Sensor

USB Interface

5-pin mini-B, Connect to PC for data transfer

Headset

2.5 mm minature phone plug

External Trigger In

BNC, female, 50 Ω, Maximum Input ± 5 VDC

Display

Size

8.4”

Resolution

800 x 600

Battery

Type

Li-Ion

Battery Operation

2.5 hours, typical

Electromagnetic Compatibility

European Union

CE Mark, EMC Directive 89/336/EEC, 92/31/EEC, 93/68/EEC and
Low Voltage Directive 73/23/EEC, 93/68/EEC

Australia and New Zealand

C-tick N274

Interference

EN 61326-1

Emissions

EN 55011

Immunity

EN 61000-4-2/-4-3/-4-4/-4-5/-4-6/-4-11

Safety

Safety Class

EN 61010-1 Class 1

Product Safety

IEC 60950-1 when used with Company supplied Power Supply

Environmental

Operating Temperature

–10 ºC to 55 ºC

Maximum Humidity

85%

Shock

MIL-PRF-28800F Class 2

Storage

–51 ºC to 71 ºC

Altitude

4600 meters, operating and non-operating

Size and Weight

Size

315 mm x 211 mm x 94 mm, (12.4 in x 8.3 in x 3.7 in)

Weight

3.1 kg, (6.9 lbs)