Options
OPT.73 Wide-range FM demodulation
Measuring amplitude range: > -50 dBm + input attenuation value
(at center frequency 1 GHz, RBW Wide,
-20 dB or more than reference level)
FM Deviation
Measuring range
2.5 MHz, 1 MHz, 500 kHz, 250 kHz, 100 kHz,
50 kHz, 25 kHz, 10 kHz
Linearity error*:
(2 % of measuring range)
Offset error*:
(4 % of measuring range + K + Readout of
frequency x Frequency reference accuracy)
K; 8 kHz (measuring range 2.5 MHz to 250 kHz)
2 kHz (measuring range 100 kHz to 10 kHz)
Demodulation frequency
bandwidth (3 dB):
300 kHz (nominal)
* These errors are values obtained by executing “FM Demod ALL CAL” software,
after warming up the R3132/3132N/3162 for 30 minutes or more.
OPT.74 Tracking generator
Frequency range:
100 kHz to 3.0 GHz
Output level range:
0 to -59.9 dBm
Output level accuracy:
0.5 dB (30 MHz, -10 dBm, 20 to 30
°
C)
Output level flatness:
1.0 dB (100 kHz to 1 GHz)
1.5 dB (100 kHz to 3 GHz)
(-10 dBm, 30 MHz reference)
Spurious
Harmonics:
-20 dBc (output level = -10 dBm)
Non-harmonics:
-30 dBc (output level = -10 dBm)
R3132/3162-9E Jan. '02