Specifications – Atec Agilent-11970 Series User Manual
Page 3
3
Input frequency, GHz
Nominal frequency response & conversion loss
18
26.5
33
40 50 60
75
110
Conversion loss, dB
50
45
40
35
30
25
20
K
A
Q
U
V
W
Specifications
Specifications describe the device’s warranted performance over the temperature range 0 to 55° C (except where noted).
IF range
DC to 1300 MHz
LO amplitude range
+14 to +18 dBm
1
Calibration accuracy
(with IF of 321.4 MHz or 322.5 MHz)
11970A/Q/U
11970V/W
±2.0 dB with LO amplitude range of +14.5 to +16 dBm;
±3.0 dB with LO amplitude range of +16 to +18 dBm
±2.6 dB with LO amplitude range of +14.5 to +16 dBm;
±3.2 dB with LO amplitude range of +16 to +18 dBm
Nominal RF input SWR
11970A/Q/U
11970V/W
< 2.2:1
< 2.6:1
Bias requirements
None
Nominal odd-order suppression
11970A/Q/U
11970V/W
> 20 dB
> 15 dB
Maximum CW RF input level
+20 dBm (100 mW)
Maximum peak pulse power
+24 dBm (250 mW) with < 1 μs
pulse (average power +20 dBm)
11970 Series tested with Agilent PSA spectrum analyzers
Agilent model
number
Frequency
range
(GHz)
LO
harmonic
number
Maximum
conversion
loss (dB)
Nominal spectrum
analyzer noise (dBm)
1 kHz BW
Frequency
response
(dB)
Nominal gain
compression
(dBm)
11970K
18 to 26.5
6–
24
–105
±1.9
–3
11970A
26.5 to 40
8–
26
–102
±1.9
–5
11970Q
33 to 50
10–
28
–101
±1.9
–7
11970U
40 to 60
10–
28
–101
±1.9
–7
11970V
50 to 75
14–
40
–92
±2.1
–3
11970W
75 to 110
18–
46
–85
±3.0
–1
Nominal performance, as noted in the following tables, while not warranted, provides information useful in applying the device.
1. An LO power of between 14.5 to 16 dBm at the mixer’s LO input is necessary to achieve the given frequency response and spectrum analyzer amplitude
accuracy specifications. When LO power varies between 14 to 18 dBm at the mixer’s LO input, add ±1 dB to the frequency response and spectrum analyzer
amplitude accuracy specification.