Accuracy

Instrumentation

Please add the corresponding power sensor linearity per-
centage; see Tables 6a and 6b for the E9320 sensors.

Average only mode:
Absolute

Logarithmic: ±0.02 dB
Linear: ±0.5%

Relative

Logarithmic: ±0.04 dB
Linear: ±1.0%

Normal mode:

Time Base Accuracy

0.01%

1 mW power reference

Power output:

1.00 mW (0.0 dBm). Factory
set to ±0.4% traceable to the
National Physical Laboratories
(NPL), UK

2

Accuracy:

For two years

±0.5% (23 ± 3 °C)
±0.6% (25 ± 10 °C)
±0.9% (0 to 55 °C)

Frequency:

50 MHz nominal

SWR:

1.06 maximum (1.08 maximum
for Option E41xA-003)

Connector type:

Type N (f), 50 ohms

Measurement characteristics:

Measurements:

Average power
Peak power
Peak-to-average ratio
Measurements between two time
offsets (time-gating)

Averaging:

Averaging over 1 to 1024 readings
is available for reducing noise

Measurement speed (GPIB)

Over the GPIB, three measurement speeds are available
(normal, x 2 and fast). The typical maximum speed is
shown in the table below.

Table 2. Measurement speed for different sensor types

Channel functions

A, B, A/B, B/A, A-B, B-A and
Relative

Storage registers

10 instrument states can be
saved via the Save/Recall menu.

Predefined setups

For common wireless standards (GSM900, EDGE, NADC,
iDEN, Bluetooth, IS-95 CDMA, W-CDMA and cdma2000),
predefined setups are provided.

Calibration temperature

1

Temperature

±5 °C

0 to 55 °C

Absolute accuracy (log)

±0.04 dB

±0.08 dB

Absolute accuracy (linear) ±0.8%

±1.7%

Relative accuracy (log)

±0.08 dB

±0.16 dB

Relative accuracy (linear)

±1.6%

±3.4%

1. Power meter is within ±5 °C of its calibration temperature.
2. National metrology institutes of member states of the Metre Convention, such as the

National Institute of Standards and Technology in the USA, are signatories to the
ComitÈ International des Poids et Mesures Mutual Recognition Arrangement. Further
information is available from the Bureau International des Poids et Mesures, at
http://www.bipm.fr/

3. Fast speed is not available for 8480 and N8480 series sensors.
4. Maximum measurement speed is obtained by using binary output in free run trigger.
5. For E9320 sensors, maximum speed is achieved using binary output in free run

acquisition.

Sensor type

Measurement speed

(readings/second)

Normal x 2 Fast

3,4

E-Series E9320 Average only mode

20

40

400

peak and average
sensors

Normal mode

5

20

40 1000

E-Series CW and E9300 average power

20

40

400

sensors

8480 and N8480 Series sensor

20

40

N.A.