Range, Zero set (internal) zero set (external) zero drift, Measurement noise – Atec Agilent-U2000 Series User Manual

Page 13: Noise per sample

background image

13

Normal Mode

Range

1

Zero set
(internal)

Zero set
(external)

Zero drift

2

Measurement
noise

3

Noise per
sample

4

U2000/1/2A

–38 to –15 dBm
–20 to –6.5 dBm
–7.5 to –2 dBm
–4 to 15 dBm
10 to 20 dBm

47 nW
530 nW
30 µW
32 µW
270 µW

43 nW
480 nW
27 µW
30 µW
200 µW

25 nW
230 nW
19 µW
24 µW
110 µW

28 nW
300 nW
20 µW
21 µW
180 µW

90 nW
1 µW
55 µW
85 µW
550 µW

U2000/1/2H

–28 to –5 dBm
–10 to 3.5 dBm
–2.5 to 8 dBm
8 to 25 dBm
20 to 30 dBm

730 nW
5.3 µW
330 µW
440 µW
3.9 µW

500 nW
4.8 µW
270 µW
300 µW
2.8 mW

300 nW
3 µW
190 µW
300 µW
1.1 mW

310 nW
5 µW
230 µW
260 µW
2.8 mW

900 nW
10 µW
550 µW
850 µW
5.5 mW

U2000/1B

–8 to 15 dBm
10 to 23.5 dBm
22.5 to 28 dBm
26 to 44 dBm

47 nW
530 nW
30 µW
32 µW

43 µW
480 µW
27 mW
34 mW

25 µW
230 µW
19 mW
24 mW

28 µW
300 µW
20 mW
21 mW

90 µW
1 mW
55 mW
85 mW

1. Condition: (i) 0 to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.

2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.

3. The number of averages at 1 for Normal speed, gate length of 2.27 ms, measured over one minute interval and two standard deviations.

4. The Noise Per Sample specification is only applicable for gated power working range stated in the “Normal Mode Key Specifications and Characteristics” table.