P-Series Wideband Power Sensor Specifications

The P-Series wideband power sensors are designed for use with the P-Series power meters only.

Sensor
model

Frequency
range

Dynamic range

Damage level

Connector
type

N1921A

50 MHz to
18 GHz

–35 dBm to +20 dBm (≥ 500 MHz)
–30 dBm to +20 dBm (50 MHz to 500 MHz)

+23 dBm (average power);
+30 dBm (< 1 μs duration) (peak power)

Type N (m)

N1922A

50 MHz to
40 GHz

–35 dBm to +20 dBm (≥ 500 MHz)
–30 dBm to +20 dBm (50 MHz to 500 MHz)

+23 dBm (average power);
+30 dBm (< 1 μs duration) (peak power)

2.4 mm (m)

1. Beyond 70 % humidity, an additional 0.6 % should be added to these values.

Maximum SWR

Frequency band

N1921A

N1922A

50 MHz to 10 GHz

1.2

1.2

10 GHz to 18 GHz

1.26

1.26

18 GHz to 26.5 GHz

1.3

26.5 GHz to 40 GHz

1.5

Sensor Calibration Uncertainty

1

Definition: Uncertainty resulting from non-linearity in the sensor detection and
correction process. This can be considered as a combination of traditional lin-
earity, cal factor and temperature specifications and the uncertainty associated
with the internal calibration process.

Frequency band

N1921A

N1922A

50 MHz to 500 MHz

4.5 %

4.3 %

500 MHz to 1 GHz

4.0 %

4.2 %

1 GHz to 10 GHz

4.0 %

4.4 %

10 GHz to 18 GHz

5.0 %

4.7 %

18 GHz to 26.5 GHz

5.9 %

26.5 GHz to 40 GHz

6.0 %

Physical characteristics

Dimensions

N1921A

N1922A

135 mm x 40 mm x 27 mm
(5.3 in x 1.6 in x 1.1 in)
127 mm x 40 mm x 27 mm
(5.0 in x 1.6 in x 1.1 in)

Weights with cable

Option 105
Option 106
Option 107

0.4 kg (0.88 Ib)
0.6 kg (1.32 Ib)
1.4 kg (3.01 Ib)

Fixed sensor cable lengths

Option 105
Option 106
Option 107

1.5 m (5 feet)
3.0 m (10 feet)
10 m (31 feet)