Environmental specifi cations, Other environmental information – Atec Agilent-83017A-83020A User Manual
Page 6
![background image](/manuals/318540/6/background.png)
6
Product specifi cations (continued)
Model
number
83006A 83017A 83018A
*Bias voltage and current
(nominal) mA
12 ±1 Vdc at 410 ±85 mA
12 ±1 Vdc at 780 ±140 mA
12 ±1 Vdc at 1940 ±123 mA
–12 ±1 Vdc at 10 ±5 mA
–12 ±1 Vdc at 20 ±2 mA
–12 ±1 Vdc at 10 ±5 mA
RF connectors
3.5 mm (f)
3.5 mm (f)
3.5 mm (f)
Detector output
N/A
BNC (f)
BNC (f)
Detector sensitivity
N/A
15 μV/μW
4 μV/μW
Detector polarity
N/A Negative
Negative
Weight: net shipping
0.64 kg (1.4 lb)
0.64 kg (1.4 lb)
1.8 kg (4.0 lb)
1.32 kg (2.9 lb)
1.32 kg (2.9 lb)
2.9 kg (6.4 lb)
*Do not apply positive voltage before negative voltage.
Environmental specifi cations
Temperature coeffi cient
of gain
–0.07 dB/° C
–0.1 dB/° C
–0.13 dB/° C
Operating temperature
0 to +55° C
0 to +55° C
0 to +55° C
Storage temperature
–40 to +70° C
–40 to +70° C
–40 to +70° C
Other environmental information
EMC
1
IEC 61326:1997/EN 61326:1997
CISPR 11:1997/EN 55011:1998, Group 1, Class A
Safety
IEC 348:1978/HD 401 S1:1981
CAN/CSA-C22.2 No. 231 (Series M-89)
Moisture resistance
65° C at 95% RH for 10 days per Mil-Std-883C method 1004.5
Random vibration
5.2 G (rms) to 2000 Hz per Mil-Std-883C method 2026 test condition 11A
Shock
1500 G (peak), 0.5 ms per Mil-Std-883C method 2002.3 test condition B
Altitude, non-operating
15,000 m per Mil-Std-883C method 1001 test condition C
1. This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.