beautypg.com

Environmental specifi cations, Other environmental information – Atec Agilent-83017A-83020A User Manual

Page 6

background image

6

Product specifi cations (continued)

Model

number

83006A 83017A 83018A

*Bias voltage and current
(nominal) mA

12 ±1 Vdc at 410 ±85 mA

12 ±1 Vdc at 780 ±140 mA

12 ±1 Vdc at 1940 ±123 mA

–12 ±1 Vdc at 10 ±5 mA

–12 ±1 Vdc at 20 ±2 mA

–12 ±1 Vdc at 10 ±5 mA

RF connectors

3.5 mm (f)

3.5 mm (f)

3.5 mm (f)

Detector output

N/A

BNC (f)

BNC (f)

Detector sensitivity

N/A

15 μV/μW

4 μV/μW

Detector polarity

N/A Negative

Negative

Weight: net shipping

0.64 kg (1.4 lb)

0.64 kg (1.4 lb)

1.8 kg (4.0 lb)

1.32 kg (2.9 lb)

1.32 kg (2.9 lb)

2.9 kg (6.4 lb)

*Do not apply positive voltage before negative voltage.

Environmental specifi cations

Temperature coeffi cient
of gain

–0.07 dB/° C

–0.1 dB/° C

–0.13 dB/° C

Operating temperature

0 to +55° C

0 to +55° C

0 to +55° C

Storage temperature

–40 to +70° C

–40 to +70° C

–40 to +70° C

Other environmental information

EMC

1

IEC 61326:1997/EN 61326:1997

CISPR 11:1997/EN 55011:1998, Group 1, Class A

Safety

IEC 348:1978/HD 401 S1:1981

CAN/CSA-C22.2 No. 231 (Series M-89)

Moisture resistance

65° C at 95% RH for 10 days per Mil-Std-883C method 1004.5

Random vibration

5.2 G (rms) to 2000 Hz per Mil-Std-883C method 2026 test condition 11A

Shock

1500 G (peak), 0.5 ms per Mil-Std-883C method 2002.3 test condition B

Altitude, non-operating

15,000 m per Mil-Std-883C method 1001 test condition C

1. This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.